![NF FRA5022 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/nf/fra5022/fra5022_instruction-manual_1684892040.webp)
3.3 I/O Terminals
FRA5022
3-10
3.3.2 Analyzer Input Terminal
The main specifications for the analyzer input CH1 and CH2 are as bellow.
Input impedance: 1M
Ω
, parallel 60pF
Measurement voltage range:
±
10V
Non-destructive maximum input voltage:
±
24V
Do not apply a signal that exceeds the non-destructive maximum input
voltage.
Otherwise,
FRA5022
may be damaged.
■
Input when the power is turned off
When the device is powered off or waiting to be ready for measurement after power-on,
the input terminals as well as the external conductor (ground) are isolated from the
internal parts.
Figure 3-4 Analyzer input terminal
■
CH1 input for CH2/OSC
When the measurement mode is set to CH2/OSC, CH1 input terminal is open.
The analyzer CH1 input is connected internally to the oscillator output.
■
Operation on excessive input
When the input signal exceeds the measurement range (
±
10V) or when the excessive
input voltage specifed for each channel is exceeded, the OVER lamp for the relevant
channel illuminates. Depending on the settings for operations with excessive input, a
sound may be generated. The OVER lamp or warning sound will not be automatically
cleared even after the excessive input is removed. Clear by pressing the ENTR/RESET
ERROR key.
1M
Ω
INPUT
CH1 CH2
from
OSCILLATOR
!
CAUTION
Содержание FRA5022
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Страница 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Страница 30: ...FRA5022 2 12...
Страница 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Страница 119: ...FRA5022 5 41...
Страница 131: ...FRA5022 7 8...
Страница 141: ......
Страница 143: ......
Страница 144: ......