![NF FRA5022 Скачать руководство пользователя страница 110](http://html1.mh-extra.com/html/nf/fra5022/fra5022_instruction-manual_1684892110.webp)
5.5 Command Explanation
FRA5022
5-32
SYSTem:ERRor?
Description:
Queries the error status
Parameters:
None
Response:
Error number, error message
Error number: format
<
NR1
>
, range
-
32768 to +32767
Error message: format
<
SRD
>
Details
・・・
“6.1.3 Remote Control Errors”
Query example:
SYST:ERR?
Response example:
0,”No error”
SYSTem:OVERload:RELease
Description:
Releases the error just like RESET ERROR key from the front
panel.
Example of released error
・
Input signal exceeds measurement possible voltage range
・
User set excessive input detection level is exceeded
・
Output overload
Parameters:
None
Example:
SYST:OVER:REL