2.2 Installation
FRA5022
2-4
2.2.3 Rack Mounting
FRA5022
can be mounted on a 19-inch IEC rack, an EIA specification rack or a JIS
standard rack by the use of a rack-mount adapter (optional).
First, mount the rack-mount adapter on the device as shown in
“Fig. 2-3 Mounting
rack-mount adapter (mm-
rack)”
or
“Fig. 2-4 Mounting rack-mount adapter (inch-rack)”
, and
then, mount the device on the rack.
Take the following precautions when you mount the device on the rack:
•
Support the
FRA5022
by all means by installing some supports such as
rails on the rack.
Otherwise,
FRA5022
may fall, causing injury or damaging itself.
•
Make sure that the rack is sufficiently air-cooled by providing appropriate
ventilation ports or cooling fans.
FRA5022
has air inlets at the bottom panel. When placing it on top of
another device, allow a clearance of 20mm or more under
FRA5022
.
Otherwise, performance degradation or damage may be incurred due to
the device temperature exceeding the specified level.
Remove the feet as shown in
“Fig. 2-5 Removal of feet”
when they interfere with the rail.
!
CAUTION
Содержание FRA5022
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Страница 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Страница 30: ...FRA5022 2 12...
Страница 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Страница 119: ...FRA5022 5 41...
Страница 131: ...FRA5022 7 8...
Страница 141: ......
Страница 143: ......
Страница 144: ......