6.1 Error Messages
FRA5022
6-2
6.1
Error Messages
Self-diagnosis is performed when the power is turned on. If an error is detected, an error
message is displayed in the lower part of the start screen. Similarly, when improper
values are being set via the remote control interface, an error message is displayed. Error
messages are also displayed when your operation via the front panel has some
restrictions.
Error messages and their causes are described below, together with required actions.
Some other messages may be displayed depending on the situation.
6.1.1 Errors at Power-On
Error messages
Cause
Required action
ROM ERROR
Internal ROM error
Turn the power off, wait for at
least 3 seconds and turn the
power on again. If the error
message persists after turning
the power on several times, it
will be necessary to make
repairs. Contact NF
Corporation or one of our
representatives.
SDRAM ERROR
Internal SDRAM error
BACKUP MEMORY
ERROR
Battery backup contents are
broken.
・
Backup battery is empty
・
Backup memory fault
・
Temporary internal data error
due to power-off, etc. while
operating on the memory
ANALYZER ERROR
Analyzer malfunction
CALIBLATION
ERROR
Self calibration error
Problem with the signal
measurement system in the
oscillator or the analyzer.
When an error message is displayed, check
“Installation Conditions”
in
“2.2 Installation”
and
verify that all the operation conditions are met.
When ROM or SDRAM error is detected, further operation is not possible.
When battery backup contents are broken, the device starts after settings and
measurement data memory are initialized. When that happens, the line after the above
error message indicates which memory category has been initialized.
STn:
Settings memory n = 0..9
NS:
Settings not indicated on the left
SA:
Data memory A for sweep measurements
SB:
Data memory B for sweep measurements
PA:
Data memory A for spot measurements
PB:
Data memory B for spot measurements
When signal system error during analyzer self check or self calibration is found, error
messages are displayed and further operation is not possible.
Содержание FRA5022
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Страница 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Страница 30: ...FRA5022 2 12...
Страница 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Страница 119: ...FRA5022 5 41...
Страница 131: ...FRA5022 7 8...
Страница 141: ......
Страница 143: ......
Страница 144: ......