5.1 Preparations Before Use
FRA5022
5-2
5.1
Preparations Before Use
FRA5022
can be remote controlled by GPIB or USB-TMC (TMC: Test and Measurement
Class). By sending program messages from the controller it is possible to control the
device similarly to panel operations and to receive measurement values or setting status
as response messages.
・
The interface assumes that GPIB or USB is used in a favorable environment. Avoid
using them in a noisy environment.
5.1.1 USB Preparations
Install the USB-TMC class driver to the controlling computer and connect with a USB
cable which is sold commercially. It is possible to download the driver install file from
National Instruments Corporation's web site. The driver installation process is described
below.
1) Go to the National Instruments Corporation web site and search for the VISA
Run-time Engine page.
2) From VISA Run-time Engine's page, download VISA Run-time Engine. User
registration is required. Download VISA Run-time Engine Ver3.3 or superior.
3) The downloaded file is a self extracting archive. Install after extraction.
4) After the process is correctly completed, the USB-TMC class driver is installed.
The above information is effective at the time of the creation of this instruction manual.
For more details, see the National Instruments Corporation's web site. The web site
address or provided services can be modified without prior notice.
5.1.2 GPIB Preparations
Install a GPIB card/controller board on the controlling computer and connect the
board/controller card to
FRA5022
using a GPIB cable commercially available. For details,
see the operation manual of the GPIB card/controller board.
5.1.3 Remote Control Interface Selection
In
FRA5022
it is possible to select GPIB or USB as remote control interface. It is not
possible to use GPIB and USB simultaneously. Select either one from the system menu
screen. First, press the MENU key to display the menu screen. Use the left/right cursor
keys
△
▽
to select the <SYSTEM> tab. Use the up/down cursor keys to select the
following items and select either GPIB or USB.
Items:
INTERFACE connect
Value range: 0) GPIB 1) USB
Содержание FRA5022
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Страница 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Страница 30: ...FRA5022 2 12...
Страница 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Страница 119: ...FRA5022 5 41...
Страница 131: ...FRA5022 7 8...
Страница 141: ......
Страница 143: ......
Страница 144: ......