5.3 Explanation of Each Program Message
BP4620
5-16
5.3.4 Measurement
related
MACC?
Description:
Returns the AC current measurement value.
Response format:
0.0 Ap-p to 80.0 Ap-p (returns 99.9 Ap-p in the case of measurement
over-range)
Caution:
Measurement is performed in units of 0.1 Ap-p.
MACV?
Description:
Returns the AC voltage measurement value.
Response format:
0 Vp-p to 400 Vp-p (999 Vp-p is returned in the case of measurement
over-range)
Caution:
Measurement is performed in units of 1 Vp-p.
MCUR?
Description:
Returns the DC current measurement value.
Response format:
40.00 A to 40.00 A (99.99 A is returned in the case of measurement
over-range)
Caution:
Measurement is performed in units of 0.01 A.
MVLT?
Description:
Returns to the DC voltage measurement value.
Response format:
200.0 V to 200.0 V (999.9 V is returned in the case of measurement
over-range)
Caution:
Measurement is performed in units of 0.1 V.
5.3.5
Arbitrary waveform memory
The settings performed with ARBM and ARBT are done for the arbitrary waveform memory specified
with ARBN. If no specification is made with ARBN, arbitrary waveform memory 2 is used as the
default. To control the arbitrary waveform saved to another memory, first specify the memory number
with ARBN.
ARBM
Description: Controls
arbitrary waveform memory.
Parameter: 0:
Clear
Setting example:
Clears arbitrary waveform.
ARBM
0
Содержание BP4620
Страница 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00016125 006 BIPOLAR DC POWER SUPPLY BP4620 Instruction Manual...
Страница 3: ......
Страница 15: ......
Страница 16: ...BP4620 1 1 1 OVERVIEW 1 1 General 1 2 1 2 Features 1 2 1 3 Principle of Operation 1 4...
Страница 21: ......
Страница 39: ......
Страница 165: ...6 1 Error Messages BP4620 6 4 Figure 6 1 Protective Function Related Error Display Examples...
Страница 171: ......
Страница 187: ......
Страница 206: ...BP4620 9 1 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 2...
Страница 211: ......
Страница 213: ......