Before you insert tubing into the
PUMP CHANNEL
, verify that the
PUMP
is in the
OFF
control mode.
With reference to Figure 33 below
, follow these steps to properly load
tubing into the
PUMP HEAD
:
Figure 33: Loading Pump Tubing
1.
Open the
PUMP
cover to gain access to the interior of the pump.
2.
Select the desired tubing size (
see Table 6 in Section 11.4 for reference
) and cut a
length sufficient to reach from the inlet source, through the pump, and to the outlet
recipient, allowing a few extra inches.
3.
Form a loop large enough to go around the pump head.
PUMP HEAD ROTOR
UPPER
SPRING-LOADED
CLAMP
LOWER
SPRING-LOADED
CLAMP
PUMP CHANNEL
PUMP COVER
PUMP HEAD