112
BioFlo/CelliGen 115
y
M1369-0050
User’s Guide
16.3
Vessel & Tubing
After each and every run, clean the vessel and the headplate with its associated parts.
All tubing and filters should be replaced.
16.4 Periodic
Inspection
At three-month intervals
, perform the following checks and inspections.
#
NOTE:
Before you begin, make sure that the power switch is in the OFF position
and that the power supply has been disconnected.
1.
Check all controls and accessible items (power switch, connectors, screws, nuts
and bolts) to make sure they are properly tightened. Tighten any loose item(s).
2.
Check that all controls and connectors are free of dust.
3.
Check that all O-rings in the headplate and impellers are intact and in good
condition. Replace those that are not.
16.5
Agitator Bearing Housing
Every 3-6 months, the ball bearings and the shaft seals in the bearing housing
should be checked and cleaned. Replace any worn-out bearings and/or shaft
seals.
16.5.1
Motor Assembly Replacement
WARNING!
NO ONE BUT A PROFESSIONAL SERVICE PERSON should touch
electric or electronic parts or assemblies in the control cabinet.
If the motor assembly should require replacement, call for an authorized NBS
service technician.