Page
16
of
76
3.3.4 Digital Input / Output
DSUB-9 Female connector for digital input and output
The digital input is isolated and can measure up to +50 VDC with programmable Schmitt trigger
levels. Digital output is also isolated using a PhotoMOS relay.
3.3.5 Communication Ports
DSUB-9 Male connector for CAN 2.0B and RS-485 communication
This interfaces can be used for additional sensor connection (pyranometer, weather,
temperature etc.). All data are fully synchronized to the analoge input data (voltage, current,
additional AI). It 12VDC for power supply (if needed).
Pin
Signal
Description
1
DO0
Isolated digital output 0
2
DO1
Isolated digital output 1
3
+12V
+12V power source
4
DI0
Isolated digital input 0
5
DI1
Isolated digital input 1
6
DO0_COM
Return path for output 0
7
DO1_COM
Return path for output 1
8
GND
Power ground (for +12V power)
9
DI_GND
Isolated digital input ground
Pin
Signal
Description
1
RS-485_0B
RS-485 B(+)
2
485_GND
GND for RS-485
3
+12V
+12V power source
4
CANL0
CAN low
5
CAN_GND
GND for CAN2.0B
6
RS-485_0A
RS-485 A(-)
7
NC
Not Connected
8
GND
Power ground
9
CANH0
CAN high
Содержание PQA8000