Page
14
of
76
10pin LEMO connector
Current sensors like Clamps, Rogowski-coils and other current sensors can be connected
using the following pin assignment.
Depending on the hardware the channels are assigned to different AI channels.
Pin
Signal
PQA8000H
1
2
Signal-
Signal- / GND
3
FGND
FGND
4
NC
Ro
5
TEDS
TEDS
6
GND
GND
7
+3.3V
Isolation GND
8
+12V
Iso9V
9
+15V
+15V
10
-15V
-15V
AI
Channel
Name
AI4
I1
AI5
I2
AI6
I3
AI7
IN
AI8
IPE
Содержание PQA8000