54
Use only copper wire for power wiring. Power extension cable must be constructed in accordance with
applicable hospital and electrical codes. Cable should be constructed of
SJT, SJTO, SJO, ST, SO, STO
or
equivalent flexible cord per US National Difference Clause 59.1DV. Users outside of the U.S. should adhere
to the applicable electrical codes for their country.
24 Volt Extension Cables
The cable requires one each of the following connectors: Switchcraft SL403F (female) and Weidmuller BL
3.81/3. See illustrations below.
Pin
1
2
3
+ 24
VDC
GND
Shield
Switchcraft 403F connector
X
Y
Z
Switchcraft SL403F
Weidmuller BL 3.81/3
Pin 1
24 Volt Extension Cable Schematic
DC Power Connectors and Pin Outs
1
2
3
Weidmuller BL 3.81/3 connector
Pin
1
2
3
+ 24
VDC
GND
Chassis