Dedicated Ground
A dedicated isolated/insulated ground
{
XE "Dedicated ground specifications"
}
for the 5000
i
system must be provided. If possible, route a dedicated ground conductor
{
XE
"Conductor:Ground, dedicated"
}
between the main power panel and the 5000
i
system branch
circuit (with the dedicated feeder). If a dedicated ground conductor is not feasible,
Chapter 2. Scanner Installation
Page 11