Model 7600 Family PWC User’s Manual
9
1. Check container for any obvious evidence of
damage.
2. Remove contents of the shipping container.
3. Inspect the scale for shipping damage.
Immediately report any damage to the
shipper.
1. Mount the scale on a stable, level surface
free from air currents and vibration. Be sure
the scale platter does not touch any adjacent
surfaces.
2. To install the scale surface flush with a
countertop, use these dimensions to guide
construction:
Platform
Minimum Cutout
Dimensions
Dimensions
7620
14" W
14.75" W
12.5" D
13.25" D
4.2" Min Ht.
7680
18" W
18.75" W
18" D
18.75" D
4.6" Min Ht.
3. Loosen the plastic collars (7620) or jam nut
(7680) on the leveling feet. Level the scale by
using the level bubble under the scale platter
as a guide. Be sure all four feet are in firm
contact with the counter, then tighten all
collars (nuts).
Initial Setup
Unpacking the Scale
Installing the Scale