Model 7600 Family PWC User’s Manual
22
Follow these steps to access and configure the
items in the
CONFIG
menu. Refer to Figure 2.
1. From the
DIAG
display press the
ZERO
key,
or from normal weighing mode, move Switch
1 to Menu mode or OPEN position, then
press the
ZERO
key…
CONFIG
is displayed.
2. Press the
UNITS
key…
FILTER
is displayed.
3. Press the
UNITS
key…
The current setting is displayed. Use the
ZERO
key to toggle between
FLTR ON
and
FLTR OFF
4. Press the
UNITS
key…
Filter selection is stored.
5. Press the
ZERO
key…
BAUD
is displayed.
6. Press the
UNITS
key…
The current setting is displayed. Use the
ZERO
key to toggle between the five
choices:
1200, 2400, 4800, 9600
, or
19200
baud
7. Press the
UNITS
key…
Baud rate selection is stored.
8. Press the
ZERO
key…
PARITY
is displayed.
9. Press the
UNITS
key…
The current setting is displayed. Use the
ZERO
key to toggle between the three
choices:
EVEN, ODD, NONE.
Step-by-Step
Instructions for
CONFIG Mode
Press the
ZERO
key to
scroll through lists of
selections.
Press the
UNITS
key
to make a selection
Tip: Quickly and easily
view current scale
configuration directly
from the front panel
without opening the
scale or setting
switches as follows:
During the display
segment test on
power-up, press the
UNITS
key. The
display will prompt
ABORT
followed
by
BAUD
. Press the
ZERO
key to scroll
through the choices, or
press the
UNITS
key to view a current
scale configuration.
When you are done,
press the
ZERO
key
until
DONE
is dis-
played. Press the
UNITS
key to return to
the normal weighing
mode.