Model 7600 Family PWC User’s Manual
25
5. Press the
ZERO
key to toggle between the
choices
UNIT ON
or
UNIT OFF.
Once your
choice is displayed, press the
UNITS
key…
See above for the definitions of calibrat-
ing the scale using
UNIT ON
or
UNIT
OFF.
6. Press the
ZERO
key to toggle between the
choices. When the choice you want is
displayed, press the
UNITS
key…
The scale prompts
RNG ON, INT ON,
or
RNG OFF
. See Table 5 in
7600 Menu
Structure - Glossary,
for definitions of
multi-range functions.
7. Press the
ZERO
key to toggle between
choices.
8. Press the
UNITS
key to accept…
The scale then prompts
LOAD O
.
9. Clear all weight from the scale platter and
press the
UNITS
key…
After a brief wait
LOAD 100
(span
weight) is displayed. Alternate calibration
points can be chosen using the
ZERO
key to toggle between choices. See
Table 6 in
7600 Menu Structure - Glos-
sary,
10. Place chosen (alternate) calibration weight
on the scale and press the
UNITS
key…
After a brief wait,
DONE
is displayed.
The scale then displays
CAL
.
11. Remove the calibration weight and return
Switch 1 to the closed position…
The scale returns to normal weighing
mode.
The scale is now tested, configured and cali-
brated. It is ready for use in your application.
Close
Switch 1
or
unplug scale NOW if
you don’t have correct
calibration weights.