XLTEK Protektor32
Instructions for Use
47
To set up traces to appear in the groups:
1. Click
Append
and then select a group from the menu of groups to add a trace to the test. A new
row is added to the
Traces
table.
2. To give the trace a custom name, click inside the
Name
column cell and type in a name. If you do
not type in a custom name, then EPWorks automatically names the trace after the input when you
select an input.
3. To define which inputs will create the trace, right-click the cell in the
Input
column and select an
input from the menu. (Inputs are predefined in the
Inputs
tab.)
4. The cell in the
Group
column was set when you selected a group from the
Append
button menu.
(If desired, you can change the group by right-clicking the
Group
column cell.)
5. To set the low frequency filter to filter out interference below a certain value, right-click the cell in
the
LFF
column and select a setting from the menu. The range of values for LFF is 0.1 Hz to 200
kHz.
6. The high frequency filter sets a circuit to pass all frequencies above a designated cut-off
frequency. Frequencies below the cut-off frequency are rejected or attenuated. Right-click the cell
in the
HFF
column and select a setting from the menu. The range of values for HFF is 15 Hz to
15 kHz.
7. The Notch filter minimizes and virtually eliminates the interference from nearby electrical
equipment. Right-click the cell in the
Notch
column to set the Notch filter to Off, 60 Hz or 50 Hz.
8. The sensitivity setting adjusts the sensitivity, or gain, of the channels. Right-click the cell in the
Sensitivity
column and select a setting from the menu.
9. The timebase setting adjusts the display and speed of the recording on screen. Right-click a cell
in the
Timebase
column and select a setting from the menu.
10. Right-click the cell in the
Color
column to set the color of the trace when it is not active (currently
acquiring data). Red is the default color for active traces.
11. By default,
Chime
is set to
<none>
. Right-click the cell in the
Chime
column to select a sound
from the
Chime
menu to set a sound for a specific trace.
12. When you are finished setting up traces in the
Traces
tab, click
Save
.
13. If you have set a timeline in the
Groups
tab to interleaving, you should now set the stimulation
delay. To do so, click the
Groups
tab, then click the
Calculate Stim Delay
button.
14. Click
Save
.
K) Markers Tab
Markers and marker definitions and calculations are set up in the
Markers
tab in the
Edit Test
window.
When defined, the corresponding Marker buttons appear in the toolbar window next to the
Group
menu.
Note that:
•
Only one marker of a given type can be placed on a particular set.
•
Markers can be set to mark waveforms and to calculate latency and amplitude values.
•
Markers can be set for each defined group.