XLTEK Protektor32
Instructions for Use
11
Do not install any other software than the EPWorks Software. Natus assumes no
responsibility when the equipment is not used as described in this manual.
Never touch the patient while connecting or disconnecting any connectors.
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
Protektor32 System, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
General Cautions
It is recommended that the stimulator probe be disinfected between patients with 70%
isopropyl alcohol. This is not a method of sterilization if the stimulator is used invasively.
Turn off all system power and disconnect the power cord from the system and the wall
before attempting to clean the unit. The Protektor32 headbox can be wiped clean with a
soft, damp cloth using non-conductive distilled water, electrically non-conductive inert
surfactants or an
Xltek
-approved cold sterilizing agent. It is important to dry off the units
quickly. Avoid letting liquid seep into any of the internal electronics of the system. Do not
use any abrasive cleaner on the system.
Inspect all cables and connections (especially the power cord) often for signs of fraying or
other damage. Do not operate the Protektor32 headbox if you suspect damage to any of the
cables or the power cord.
Do not leave any cables attached to the Protektor32 headbox when transporting the unit.
This may cause connections to become loose or malfunction during operation of the unit.
Do not turn on the power to the Protektor32 headbox immediately after bringing it in from a
cold environment to one at room temperature. Allow the unit to assume the ambient
environmental temperature (i.e. one hour warm up).
The isolated switch mode power supply is intended for connection only to a 110V, 120V,
220V or 240V external (wall) outlet.
Audio/Visual Stimulation Warnings
Only the Telephonics Model TDH-39P headphones and ear inserts (TIPS) provided have
been approved for use with your Protektor32. Patient isolation in accordance with
IEC60601-1
is dependent on the approved parts.