XLTEK Protektor32
Instructions for Use
41
7. To set the auto squelch or speaker so as not to respond below a certain level, and avoid 60 Hz
hum, right-click the cells in the
Squelch
column and set a squelch threshold.
8. Click
Save
.
F) Electric Stimulators Tab
To set up the Electric Stimulators:
1. Click the
Electric Stimulators
tab.
2. To add the stimulators you want, click the
Add
button and select stimulator from the drop-down
menu. Repeat until you have added all the stimulators that are needed for the test.
3. Click each cell in the
Name
column and type a name for each stimulator (e.g., “Left PTN” or
“Right PTN”).
4. To select a mode of stimulation, right-click the cells in the
Mode
column and select from single
pulse, repetitive pulse, triggered, free run and averaged. Note that the selected mode determines
which options are available in the remaining columns.
5. To set the maximum intensity of the stimulator output, right-click the cells in the
Maximum
Intensity
column and select from the available intensities (Maximum available intensity 100 mA
(milliamps)).
6. To set the duration of the stimulation, right-click the cells in the
Pulse Duration
column and
select from 0.05 ms, 0.1 ms, 0.2 ms, 0.3 ms, 0.5 ms, 1 ms.
7. To set the stimulus rate, right-click the cells in the
Pulse Rate
column, and select the desired
rate.
8. The remaining columns allow you to set
Pulse/Train
and
Train Rate
.
9. Click
Save
.