NI VirtualBench Calibration Procedure
|
© National Instruments
|
67
8.
Call
PS Measure Adjustment Point
with the following parameters:
•
Channel
=
ps/+6V
•
Reference Value
—Supplied from the DMM measurement in step 7
9.
Repeat steps 5 through 8 for each adjustment point on this channel.
10. Disconnect all connections and connect C25 V to Volts HI, and Common Floating
GND to Volts LO as shown in Figure 36. Connect GND and Common Floating GND with
a jumper.
Figure 36.
Cable Connections For +25 V Voltage Programming and Measurement
Accuracy Adjustment
11. Repeat steps 4 through 9 for the
ps/+25V
channel.
1
VirtualBench DC Power Supply
2
DMM