NI VirtualBench Calibration Procedure
Verifying Resistance
To verify the resistance function of the VirtualBench, complete the following steps:
1.
To verify resistance ≤19 kΩ, connect the red connectors on one end of the low EMF copper
cable to the banana plug connectors of the VirtualBench, and connect the connectors on the
other end of the cable to the appropriate calibrator binding posts. Figure 3 shows the correct
connections.
Figure 3.
Cable Connections for Resistance
≤
19 k
Ω
2.
Wait 2 minutes for the thermal EMF to stabilize.
Note
Thermal EMF stabilization time is not needed if the cables have not been
connected/reconnected.
3.
Call
DMM Configure Measurement
with the following parameters:
•
Function
=
Resistance
•
Auto Range
=
FALSE
•
Manual Range
=
100
4.
Set the calibrator to output 0 Ω with 2-wire compensation turned on.
5.
Call
DMM Read
to allow the calibrator to properly settle.
6.
Wait 5 seconds for settling.
7.
Call
DMM Read
20 times. Calculate and record the average value.
8.
Output 1 Ω verification point on the calibrator.
9.
Call
DMM Configure Measurement
with the following parameters:
•
Function
=
Resistance
•
Auto Range
=
FALSE
•
Manual Range
=
100
1