![National Instruments VirtualBench Скачать руководство пользователя страница 56](http://html2.mh-extra.com/html/national-instruments/virtualbench/virtualbench_calibration-procedure_3421539056.webp)
56
|
ni.com
|
NI VirtualBench Calibration Procedure
Verifying Voltage Programming and Measurement Accuracy
To verify the voltage programming and measurement accuracy on the +6 V, +25 V, and -25 V
channels, complete the following steps.
1.
Reset the DMM and configure it for the 10 V range.
2.
Connect C6 V to Volts HI, and GND to Volts LO with cabling wire, as shown in
Figure 29.
Figure 29.
Cable Connections For +6 V Voltage Programming Accuracy Verification
3.
Call
PS Initialize
.
4.
Call
PS Enable All Outputs
with the following parameter:
•
Enable Outputs
=
TRUE