42
|
ni.com
|
NI VirtualBench Calibration Procedure
24.
(VB-8034/8054)
Remove the connections from channel 3. Connect VirtualBench channel
4 to the oscilloscope calibrator active head, as shown in Figure 22.
Figure 22.
VB-8034/8054 Oscilloscope Gain Channel 4 Connection Diagram
25.
(VB-8034/8054)
Repeat steps 2 to 18 for all ranges on channel 4.
Note
(VB-8034/8054)
When calling
MSO Configure Analog Channel
in
step 4, change the
Channel
input to
mso/4
.
26.
(VB-8034/8054)
Remove the connections from channel 4.
1
VirtualBench Oscilloscope
2
Oscilloscope Calibrator
3
Oscilloscope Calibrator Active Head
2
CH 2
CH 1
1
3