NI VirtualBench Calibration Procedure
Verifying the Oscilloscope Gain
Complete the following steps to verify the VirtualBench oscilloscope gain.
1.
Connect VirtualBench channel 1 to the oscilloscope calibrator active head, as shown in
Figure 15.
Figure 15.
Oscilloscope Gain Channel 1 Connection Diagram
2.
Call
MSO Initialize
.
3.
Call
MSO Configure Timing
with the following parameters:
•
Sample Rate
=
1M
•
Acquisition Time (s)
=
0.1
•
Pretrigger Time (s)
=
1e-9
•
Sampling Mode
=
Sample
4.
Call
MSO Configure Analog Channel
with the following parameters:
•
Enable Channel
=
TRUE
•
Channel
=
mso/1
•
Vertical Coupling
=
DC
1
VirtualBench Oscilloscope
2
Oscilloscope Calibrator
3
Oscilloscope Calibrator Active Head
2
CH 2
CH 1
1
3