NI VirtualBench Calibration Procedure
|
© National Instruments
|
47
24. Repeat steps 3 to 21 for MSO channel 2.
Note
For all inputs, change the
Channel
input to
mso/2
.
25.
(VB-8012)
26.
(VB-8034/8054)
Repeat steps 3 to 21 for MSO channel 3.
Note
(VB-8034/8054)
For all inputs, change the
Channel
input to
mso/3
.
27.
(VB-8034/8054)
Repeat steps 3 to 21 for MSO channel 4.
Note
(VB-8034/8054)
For all inputs, change the
Channel
input to
mso/4
.
28. Call
MSO Close Calibration
with the following parameter:
•
Action
=
Commit
You have completed adjusting the VirtualBench oscilloscope. It is recommended that a
post-adjustment verification be performed.
Oscilloscope EEPROM Update
When an adjustment procedure for the oscilloscope is completed, the VirtualBench internal
calibration memory (EEPROM) is immediately updated.
If you do not want to perform an adjustment, you can update the calibration date and onboard
calibration temperature without making any adjustments by calling
Cal Set Calibration
Information
.
Oscilloscope Re-Verification
Repeat the
section to determine the As-Left status of the instrument.
Note
If any test fails re-verification after performing an adjustment, verify that you
have met the conditions listed in the
returning your device to NI. Refer to
for assistance in
returning the device to NI.
Function Generator (FGEN) Calibration
Procedure
This section contains the verification and adjustment procedures for the FGEN on the
VirtualBench. The calibration process includes the following steps:
1.
—Set up the test equipment.
2.
—Verify the existing operation of the instrument. This step confirms
whether the instrument is operating within its specified range prior to adjustment.