NI VirtualBench Calibration Procedure
3.
Wait 2 minutes for the thermal EMF to stabilize.
4.
Call
DMM Initialize Calibration
with the following parameter:
•
Password
=
NI
(default)
5.
Call
DMM Get Resistance Calibration Adjustment Points
. This will return
two arrays:
•
Range
—A numerical array specifying the range being calibrated
•
Adjustment Point
—A numerical array specifying the resistance values required to
perform the adjustment
6.
Output the adjustment point on the calibrator with 2-wire compensation turned on.
7.
Call
DMM Setup Resistance Calibration
with the following parameter:
•
Range
—Supplied from DMM Get Resistance Calibration Adjustment Points
8.
Wait 5 seconds for settling.
9.
Call
DMM Adjust Resistance Calibration
with the following parameter:
•
Adjustment Point
—Resistance value returned by the calibrator
10. Repeat steps 6 through 9 for the first four adjustment points.
11. To adjust resistance >19 kΩ, output the adjustment point on the calibrator with 2-wire
compensation turned off.
12. Call
DMM Setup Resistance Calibration
with the following parameter:
•
Range
—Supplied from DMM Get Resistance Calibration Adjustment Points
13. Wait 5 seconds for settling.
14. Call
DMM Adjust Resistance Calibration
with the following parameter:
•
Adjustment Point
—Resistance value returned by the calibrator
15. Repeat steps 11 through 14 for the last eight adjustment points.
16. Call
DMM Close Calibration
with the following parameter:
•
Action Input
=
Commit
17. Reset the calibrator for safety reasons.
18. Disconnect all cables from the calibrator and VirtualBench.
You have completed adjusting the resistance function of the VirtualBench.