TBX-1329 Installation Guide

Electromagnetic Compatibility

EMC/EMI............................................... CE, C-Tick and FCC Part 15

(Class A) Compliant

Electrical emissions................................ EN 55011 Class A at 10 m,

FCC Part 15A above 1 GHz

Electrical immunity................................ Evaluated to EN 61326:1998,

Table 1

Note

This device should only be operated with shielded cabling for full EMC and EMI

compliance. See the Declaration of Conformity for this product for any additional
regulatory compliance information.

Technical Support Resources

NI Web Support

National Instruments Web support is your first stop for help in solving
installation, configuration, and application problems and questions. Online
problem-solving and diagnostic resources include frequently asked
questions, knowledge bases, product-specific troubleshooting wizards,
manuals, drivers, software updates, and more. Web support is available
through the Technical Support section of

ni.com

Worldwide Support

National Instruments has offices located around the world to help address
your support needs. You can access our branch office Web sites from the
Worldwide Offices section of

ni.com

. Branch office Web sites provide

up-to-date contact information, support phone numbers, e-mail addresses,
and current events.

If you have searched the technical support resources on our Web site
and still cannot find the answers you need, contact your local office or
National Instruments corporate. For telephone support in the United States,
dial 512 795 8248. For telephone support outside the United States, contact
your local branch office:

Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20,
Brazil 011 284 5011, Canada (Calgary) 403 274 9391,
Canada (Ottawa) 613 233 5949, Canada (Québec) 514 694 8521,
China (Shanghai) 021 6555 7838, China (ShenZhen) 0755 3904939,
Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24,
Germany 089 741 31 30, Greece 30 1 42 96 427, Hong Kong 2645 3186,
India 91805275406, Israel 03 6120092, Italy 02 413091,

Содержание TBX-1329

Страница 1: ... 1125 or SCXI 1126 front connector The TBX 1329 mounts on most European standard DIN EN mounting rails The terminal block has 16 screw terminals for easy connection Two screws are provided to connect to the SCXI functional earth terminal through the SH32 32 A cable shield With the SCXI 1120 SCXI 1120D SCXI 1125 or SCXI 1126 the remaining eight pairs of screw terminals connect signals to the eight ...

Страница 2: ... a word or value that you must supply monospace Text in this font denotes text or characters that you should enter from the keyboard sections of code programming examples and syntax examples This font is also used for the proper names of disk drives paths directories programs subprograms subroutines device names functions operations variables filenames and extensions and code excerpts Safety Infor...

Страница 3: ...mode voltages you must insulate your signal wires for the highest input voltage National Instruments is not liable for any damages or injuries resulting from inadequate signal wire insulation Use only 26 14 AWG wire with a minimum voltage rating of 300 V and a temperature value of 60 C for measuring up to 300 V When connecting or disconnecting signal lines to the SCXI terminal block screw terminal...

Страница 4: ...X 1329 across the excitation outputs provided on the SCXI 1121 To reduce errors in your output voltage first you need to calculate the voltage drop across the SH32 32 A cable This voltage drop which results in an error in your measurement depends on the cable length and on the strain gauge value and configuration The SH32 32 A cable has a resistance of 0 21 Ω m To determine the amount of error int...

Страница 5: ...mpensating is to calculate the lead resistance as shown in Steps 1 through 3 then input it along with your other strain gauge parameters into the conversion formula in your National Instruments data acquisition software packages In order to minimize resistive compensation locate the load closer to your SCXI module by using shorter cable lengths or use heavy gauge wire to connect to the TBX 1329 Me...

Страница 6: ...X Each switch that switches the resistor into the circuit is labeled with SRX The X symbol represents the channel number for example SC2 is the capacitor switch for channel 2 and SR2 is the resistor switch for channel 2 Table 2 shows the TBX 1329 terminal block switch settings Table 2 TBX 1329 Terminal Block Switch Settings Switch Settings for Each Channel Description Switch Configuration SCXI 112...

Страница 7: ...ling Figure 1 TBX 1329 Parts Locator Diagram To connect the signal to the terminal block perform the following steps while referring to Figure 2 1 Remove the TBX 1329 terminal block cover by unscrewing the four captive cover screws in the cover corners These screws stay attached to the cover without falling out 1 Current Loop Receiver Resistor 2 Safety Ground Solder Lugs 3 Product Name 4 Assembly ...

Страница 8: ...ully into the terminals 4 Tighten the terminals to a torque of 5 7 in lb 56 79 Newton cm 5 Relieve strain on your signal wires by tie wrapping them to the mounting rails of your rack 6 Replace the TBX 1329 terminal block cover and tighten the captive cover screws Figure 2 TBX 1329 Parts Locator Diagram 1 Safety Ground Lugs 2 Captive Cover Screws 3 Terminal Block Connector 4 Signal Wire Entry 2 1 4...

Страница 9: ... and Terminal Block Perform the following steps to mount the SH32 32 A cable assembly and connect the TBX 1329 to your SCXI module Refer to Figures 3 through 5 as needed 1 Turn off your SCXI chassis 2 Turn off the computer that contains your data acquisition DAQ device or disconnect the device from your SCXI chassis 3 Connect the TBX cable adapter to the appropriate SCXI module and secure it by ti...

Страница 10: ...329 Installation Guide 10 ni com Figure 3 Connecting the SH32 32 A Cable to the SCXI Module 1 Backshell Mounting Screws and Ears 2 Cable Assembly 3 Thumb Screws 4 SCXI Chassis 5 TBX Cable Adapter 3 2 4 3 5 1 ...

Страница 11: ...the other end of the cable assembly to your TBX 1329 terminal block connector and secure the SH32 32 A cable by tightening both backshell mounting screws 7 Reconnect the DAQ device to your SCXI chassis 1 Backshell Mounting Screws and Ears 2 Safety Ground Lugs 3 Captive Cover Screws 4 Terminal Block Connector 5 Signal Wire Entry Note The SH32 32 A cable is not shown in the exact position for proper...

Страница 12: ... Rack Mount Installation Guide for instructions If you are not using this rack mount assembly perform the following steps to mount the TBX assembly directly onto your DIN rail 1 Snap the TBX terminal block onto the DIN rail with a firm push 2 Install the SCXI chassis using the appropriate chassis rack mount kit Note To remove the TBX terminal block from the DIN rail place a flathead screwdriver in...

Страница 13: ...250 VDC Input impedance minimum Between CH or CH terminals 2 MΩ Between CH or CH terminal and ground 1 MΩ Current receiver resistors Resistors included none Resistor sockets 8 pairs Maximum current 20 mA Maximum working voltage signal common mode Channel to ground Each channel must remain within 300 Vrms or 300 VDC of ground Channel to channel Each channel must remain within 300 Vrms or 300 VDC of...

Страница 14: ...earth ground 2 lugs Mechanical Dimensions 12 7 by 7 62 by 11 16 cm 5 by 3 by 4 4 in Weight 100 gm 3 5 oz Compatible DIN rails DIN EN 50 022 DIN EN 50 035 Environmental Operating temperature 0 to 50 C Storage temperature 20 to 70 C Relative humidity 10 to 90 noncondensing Altitude maximum 2000 m Safety Designed in accordance with IEC 61010 1 UL 3111 1 and CAN CSA C22 2 No 1010 1 for electrical meas...

Страница 15: ...echnical Support section of ni com Worldwide Support National Instruments has offices located around the world to help address your support needs You can access our branch office Web sites from the Worldwide Offices section of ni com Branch office Web sites provide up to date contact information support phone numbers e mail addresses and current events If you have searched the technical support re...

Страница 16: ...0 7625 Netherlands 0348 433466 New Zealand 09 914 0488 Norway 32 27 73 00 Poland 0 22 528 94 06 Portugal 351 1 726 9011 Singapore 2265886 Spain 91 640 0085 Sweden 08 587 895 00 Switzerland 056 200 51 51 Taiwan 02 2528 7227 United Kingdom 01635 523545 ...