Table 1. PXIe-5164 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
Power splitter Aeroflex/Weinschel
1593
Test system
characterization
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
•
Amplitude: -16 dBm to
16 dBm
•
Frequency: 50 kHz to
401 MHz
•
VSWR: <1.09
50 Ω BNC
terminator (f)
Fairview
Microwave ST3B-F
Test system
characterization
•
Amplitude: 10 dBm
•
Frequency: DC to 401 MHz
•
VSWR: <1.1
50 Ω BNC
terminator
(m)
Fairview
Microwave ST2B
Verifications:
•
RMS noise
•
Frequency: DC to 500 MHz
•
VSWR: <1.15
SMA (m)-to-
SMA (m)
cable
—
Test system
characterization
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
•
Frequency: DC to 401 MHz
•
VSWR: <1.1
•
Length: ≤1 meter
SMA (f)-to-N
(m) adapter
Fairview
Microwave
SM4226
Test system
characterization
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
•
Frequency range: DC to
401 MHz
•
VSWR: <1.05
6
|
ni.com
|
PXIe-5164 Calibration Procedure