2.
Call the niHsai Group B External Cal API v1 Host.lvclass:Passband Flatness Cal
Initialize VI with the following settings:
•
channel
: The specified channel
•
input impedance
: 1 MΩ
3.
Configure the power sensor with the following settings:
•
Power measurement: continuous average
•
Path selection: automatic
•
Averaging: automatic
•
Averaging resolution: 4 (0.001 dB)
•
Aperture: 20 ms
4.
Call the niHsai Group B External Cal API v1 Host.lvclass:Passband Flatness Cal
Configure VI to obtain the amplitude and frequency of the sine waveform to generate.
Configure the signal generator to output a sine waveform with the specified amplitude
plus
splitter loss
and frequency into a 50 Ω load.
5.
Enable the signal generator output if not already enabled.
Note
Select the splitter loss value from the list of test points from the
section for the current Test Point Frequency.
6.
Configure the power sensor to correct for the current sine wave frequency using the
power sensor frequency correction function.
7.
Wait 0.1 second for settling.
8.
Use the power sensor to measure the power in dBm. Record the result as
measured
input power
.
9.
Calculate the corrected input power using the following equation:
corrected input power
=
measured input power
+
splitter balance
Note
Select the
splitter balance
value from the list of test points from
the Test System Characterization section for the current Test Point Frequency.
10. Call the niHsai Group B External Cal API v1 Host.lvclass:Passband Flatness Cal Adjust
VI with the following settings:
•
actual frequency generated (Hz)
: The frequency of the sine waveform present on
the specified channel of the PXIe-5164
•
actual amplitude generated (dBm)
: The corrected input power of the sine
waveform present on the specified channel of the PXIe-5164
11. Repeat steps 4 through 10 until the
passband flatness cal complete
indicator from the
niHsai Group B External Cal API v1 Host.lvclass:Passband Flatness Cal Adjust VI
returns TRUE.
12. Disable the signal generator output.
Return to the main
task.
50
|