Table 1. PXIe-5164 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
DMM
NI PXI-4071
Verifications:
•
AC amplitude
accuracy
AC voltage measurement:
•
Range: 0.125 V
pk-pk
to
20 V
pk-pk
•
Input impedance: ≥10 MΩ
•
Bandwidth: ≥50 kHz
•
Accuracy at 50 kHz
–
±(0.07% of r
14 µV) for
0.125 V
pk-pk
test point
–
±(0.06% of r
71 µV) for 0.25 V
pk-pk
to 1.25 V
pk-pk
test
points
–
±(0.06% of r
707 µV) for 2.5 V
pk-pk
to 12.5 V
pk-pk
test
points
–
±(0.12% of r
35 mV) for 20.0 V
pk-pk
test point
Function
generator
NI PXI-5402/5406
or Agilent 33220A
Verifications:
•
AC amplitude
accuracy
Sine wave generation:
•
Amplitude:
–
0.125 V
pk-pk
to
2.5 V
pk-pk
into 50 Ω
–
0.125 V
pk-pk
to
20 V
pk-pk
into 1 MΩ
•
Frequency: 50 kHz
BNC Tee
(m-f-f)
Pasternack PE9174
Verifications:
•
AC amplitude
accuracy
Impedance: 50 Ω
Double
banana plug
to BNC (f)
Pasternack PE9008
Verifications:
•
AC amplitude
accuracy
Impedance: 50 Ω
4
|
ni.com
|
PXIe-5164 Calibration Procedure