19. Wait 0.1 second for settling.
20. Use the power sensor to measure the power in dBm.
21. Repeat steps 17 through 20 for each configuration in the 

Power Splitter Characterization

table, recording each result as 

splitter output 1 power

, where each configuration has a

corresponding value.

22. Calculate the splitter balance for each frequency point using the following equation:

splitter balance

 = 

splitter output 2 power

 - 

splitter output 1 power

23. Calculate the splitter loss for each frequency point using the following equation:

splitter loss

 = 

signal generator amplitude level

 - 

splitter output 2 power

24. Disconnect the 50 Ω BNC terminator (f) from splitter output 2. Refer to the remaining

assembly as the 

power sensor assembly

.

The power sensor assembly will be used in the following procedures:

Verifying 50 Ω Passband Amplitude Flatness and Bandwidth

Verifying 1 MΩ Passband Amplitude Flatness and Bandwidth

Adjusting 50 Ω Passband Amplitude Flatness and Bandwidth

Adjusting 1 MΩ Passband Amplitude Flatness and Bandwidth

Verification

The performance verification procedures assume that adequate traceable uncertainties are
available for the calibration references.

Verification of the PXIe-5164 is complete only after you have successfully completed all tests
in this section using the 

As-Found Limits

.

Refer to the following figure for the names and locations of the PXIe-5164 front panel
connectors. You can find information about the functions of these connectors in the device
getting started guide.

14

  | 

ni.com

  | 

PXIe-5164 Calibration Procedure

Содержание PXIe-5164

Страница 1: ...r Amplitude Balance and Loss 10 Verification 14 Verifying DC Accuracy 15 Verifying AC Amplitude Accuracy 19 Verifying 50 Passband Amplitude Flatness and Bandwidth 22 Verifying 1 M Passband Amplitude F...

Страница 2: ...opment environment ADE LabVIEW You can download all required software from ni com downloads Related Documentation For additional information refer to the following documents as you perform the calibra...

Страница 3: ...on Amplitude 0 9 Vpk pk into 50 Frequency 11 MHz and 99 MHz Frequency accuracy 0 25 ppm Square wave generation Amplitude 0 5 Vpk pk to 45 Vpk pk into 1 M symmetrical to ground 0 V Frequency 500 Hz Abb...

Страница 4: ...5 Vpk pk test points 0 06 of reading 707 V for 2 5 Vpk pk to 12 5 Vpk pk test points 0 12 of reading 35 mV for 20 0 Vpk pk test point Function generator NI PXI 5402 5406 or Agilent 33220A Verification...

Страница 5: ...ncy range 50 kHz to 401 MHz Power range 16 dBm to 10 dBm VSWR 1 11 Absolute accuracy 0 048 dB for 50 kHz to 100 MHz 0 063 dB for 100 MHz to 401 MHz Relative accuracy at 4 dBm 0 022 dB for 50 kHz to 10...

Страница 6: ...m characterization Amplitude 10 dBm Frequency DC to 401 MHz VSWR 1 1 50 BNC terminator m Fairview Microwave ST2B Verifications RMS noise Frequency DC to 500 MHz VSWR 1 15 SMA m to SMA m cable Test sys...

Страница 7: ...asternack PE6026 Adjustment Verifications Passband amplitude flatness and bandwidth Amplitude 10 dBm Frequency 50 kHz to 301 MHz VSWR 1 1 at 100 MHz 1 25 at 301 MHz Impedance 50 Test Conditions The fo...

Страница 8: ...ft Limits The as found limits are the published specifications for the PXIe 5164 NI uses these limits to determine whether the PXIe 5164 meets the specifications when it is received for calibration Us...

Страница 9: ...idth 3 Verification Verify the existing operation of the device This step confirms whether the device is operating within the published specification prior to adjustment 4 Adjustment Perform an extern...

Страница 10: ...e following procedures Verifying 50 Passband Amplitude Flatness and Bandwidth Verifying 1 M Passband Amplitude Flatness and Bandwidth Adjusting 50 Passband Amplitude Flatness and Bandwidth Adjusting 1...

Страница 11: ...nect the RF OUT connector of the signal generator to the input port of the power splitter using an SMA f to N m adapter and an SMA m to SMA m cable 4 Connect an SMA m to BNC m adapter to one of the po...

Страница 12: ...election automatic Averaging automatic Averaging resolution 4 0 001 dB Aperture 20 ms 9 Configure the signal generator to generate a sine waveform with the following characteristics Frequency the Test...

Страница 13: ...setup Figure 2 Connection Diagram for Measuring at Splitter Output 1 1 3 8 6 7 6 5 4 2 1 Signal generator 2 SMA f to N m adapter 3 SMA m to SMA m cable 4 Power sensor 5 BNC f to N f adapter 6 SMA m to...

Страница 14: ...the remaining assembly as the power sensor assembly The power sensor assembly will be used in the following procedures Verifying 50 Passband Amplitude Flatness and Bandwidth Verifying 1 M Passband Amp...

Страница 15: ...rifying DC Accuracy This procedure verifies the DC accuracy of the PXIe 5164 by comparing the voltage measured by the device to the value sourced by the voltage standard Caution Avoid touching the con...

Страница 16: ...113 1 38 0 83 0 15 12 1 M 0 25 V 0 0 113 1 38 0 83 0 15 13 1 M 0 25 V 5 5 113 21 4 13 0 1 5 14 1 M 0 25 V 5 5 113 21 4 13 0 1 5 15 1 M 0 5 V 0 0 225 2 61 1 75 0 23 16 1 M 0 5 V 0 0 225 2 61 1 75 0 23...

Страница 17: ...V 0 4 500 49 4 36 2 4 0 33 1 M 10 V 10 14 50 89 4 60 2 8 0 34 1 M 10 V 10 14 50 89 4 60 2 8 0 35 1 M 25 V 0 11 25 123 96 4 4 70 36 1 M 25 V 0 11 25 123 96 4 4 70 37 1 M 25 V 50 61 25 323 225 25 0 38 1...

Страница 18: ...195 5 200 0 831 552 55 0 57 1 M 5 V 197 8 200 0 816 540 48 0 58 1 M 2 5 V 198 9 200 0 808 534 48 0 1 Connect the calibrator test head to channel 0 of the PXIe 5164 2 Configure the PXIe 5164 with the...

Страница 19: ...ough 8 for each configuration listed in the DC Accuracy Verification table 12 Disable the calibrator output Verifying AC Amplitude Accuracy Follow this procedure to verify the AC amplitude accuracy of...

Страница 20: ...0 5 0 5 0 2 0 13 0 013 9 1 M 2 5 V 1 25 0 5 0 2 0 13 0 012 10 1 M 5 V 2 5 5 0 0 2 0 13 0 012 11 1 M 10 V 5 5 0 0 2 0 13 0 013 12 1 M 25 V 12 5 5 0 0 2 0 1 0 013 13 1 M 50 V 20 50 0 0 2 0 1 0 048 14 1...

Страница 21: ...the PXIe 5164 as shown in the AC Verification Test Connections figure 2 Configure the DMM with the following settings Function AC voltage Resolution 6 5 digits Min frequency 49 kHz Auto Zero Enabled A...

Страница 22: ...culate the amplitude error using the following formula AC Voltage Error 20 log10 VPXIe 5164 Measured VDMM Measured 9 Compare the amplitude error to the appropriate Limit from the AC Amplitude Accuracy...

Страница 23: ...0 29 0 117 9 1 G 0 25 V 190 1 11 0 5 0 28 0 120 10 1 G 0 25 V 210 1 11 0 5 0 27 0 124 11 1 G 0 25 V 230 1 11 0 5 0 26 0 126 12 1 G 0 25 V 245 1 11 0 5 0 25 0 127 13 1 G 0 25 V 270 1 11 0 5 0 25 0 124...

Страница 24: ...6 0 126 28 1 G 0 5 V 245 1 5 0 5 0 25 0 127 29 1 G 0 5 V 270 1 5 0 5 0 25 0 124 30 1 G 0 5 V 290 1 5 0 5 0 25 0 120 31 1 G 0 5 V 330 1 5 0 5 0 24 0 122 32 1 G 0 5 V 400 1 5 3 0 to 1 0 2 38 to 0 1 0 12...

Страница 25: ...00 1 1 3 0 to 1 0 2 38 to 0 1 0 126 49 125 M 2 5 V 0 05 9 50 1 G 2 5 V 1 1 9 0 5 0 38 0 103 51 1 G 2 5 V 10 1 9 0 5 0 38 0 103 52 1 G 2 5 V 30 1 9 0 5 0 37 0 104 53 1 G 2 5 V 60 1 9 0 5 0 35 0 105 54...

Страница 26: ...68 1 G 5 V 30 1 9 0 5 0 37 0 104 69 1 G 5 V 60 1 9 0 5 0 35 0 105 70 1 G 5 V 90 1 9 0 5 0 33 0 110 71 1 G 5 V 120 1 9 0 5 0 31 0 118 72 1 G 5 V 170 1 9 0 5 0 29 0 117 73 1 G 5 V 190 1 9 0 5 0 28 0 12...

Страница 27: ...wer sensor assembly must match the configuration used in the Test System Characterization section in which the power sensor is connected to splitter output 1 and the signal generator is connected to t...

Страница 28: ...LK IN PFI 0 AUX 0 7 2 1 3 4 3 5 1 Power Sensor 2 BNC f to N f Adapter 3 SMA m to BNC m Adapter 4 Power Splitter 5 PXIe 5164 6 SMA m to SMA m Cable 7 SMA f to N m Adapter 8 Signal Generator 2 Configure...

Страница 29: ...m Record the result as measured input power 8 Calculate the corrected input power using the following equation corrected input power measured input power splitter balance Note Select the splitter bala...

Страница 30: ...wer splitter Table 6 1 M Passband Amplitude Flatness and Bandwidth Verification Config Sample Rate S s Vertical Range Vpk pk Test Point As Found Limits dB As Left Limits dB Measurement Uncertainty dB...

Страница 31: ...G 0 5 V 30 1 5 0 7 0 57 0 109 22 1 G 0 5 V 45 1 5 0 7 0 55 0 110 23 1 G 0 5 V 60 1 5 0 7 0 53 0 110 24 1 G 0 5 V 75 1 5 0 7 0 51 0 113 25 1 G 0 5 V 90 1 5 0 7 0 5 0 111 26 1 G 0 5 V 100 1 5 0 7 0 48...

Страница 32: ...41 1 G 1 V 90 1 1 0 7 0 5 0 111 42 1 G 1 V 100 1 1 0 7 0 48 0 118 43 1 G 1 V 120 1 1 0 7 0 43 0 148 44 1 G 1 V 150 1 1 0 7 0 42 0 130 45 1 G 1 V 170 1 1 0 7 0 41 0 129 46 1 G 1 V 190 1 1 0 7 0 37 0 13...

Страница 33: ...130 61 1 G 2 5 V 170 1 9 0 7 0 41 0 129 62 1 G 2 5 V 190 1 9 0 7 0 37 0 130 63 1 G 2 5 V 200 1 9 0 7 0 33 0 136 64 1 G 2 5 V 285 1 9 3 0 to 1 0 2 07 to 0 0 152 65 125 M 5 V 0 05 9 66 1 G 5 V 1 1 9 0...

Страница 34: ...2 07 to 0 0 152 81 125 M 10 V 0 05 9 82 1 G 10 V 1 1 9 0 7 0 57 0 109 83 1 G 10 V 6 1 9 0 7 0 57 0 109 84 1 G 10 V 10 1 9 0 7 0 57 0 109 85 1 G 10 V 30 1 9 0 7 0 57 0 109 86 1 G 10 V 45 1 9 0 7 0 55...

Страница 35: ...V 30 1 9 0 7 0 57 0 109 102 1 G 25 V 45 1 9 0 7 0 55 0 110 103 1 G 25 V 60 1 9 0 7 0 53 0 110 104 1 G 25 V 75 1 9 0 7 0 51 0 113 105 1 G 25 V 90 1 9 0 7 0 5 0 111 106 1 G 25 V 100 1 9 0 7 0 48 0 118...

Страница 36: ...1 G 50 V 90 1 9 0 7 0 5 0 111 122 1 G 50 V 100 1 9 0 7 0 48 0 118 123 1 G 50 V 120 1 9 0 7 0 49 0 148 124 1 G 50 V 150 1 9 0 7 0 47 0 130 125 1 G 50 V 170 1 9 0 7 0 45 0 129 126 1 G 50 V 190 1 9 0 7...

Страница 37: ...1 9 3 0 to 1 0 2 4 to 0 0 152 1 Place the 50 SMA feed thru terminator between the splitter output and the SMA m to BNC m adapter of splitter output 2 Connect splitter output 2 of the power sensor ass...

Страница 38: ...N PFI 0 AUX 0 2 1 3 4 5 3 1 Power Sensor 2 BNC f to N f Adapter 3 SMA m to BNC m Adapter 4 Power Splitter 5 50 SMA Feed Thru Terminator 6 PXIe 5164 7 SMA m to SMA m Cable 8 SMA f to N m Adapter 9 Sign...

Страница 39: ...nction 6 Wait 0 1 second for settling 7 Use the power sensor to measure the power in dBm Record the result as measured input power 8 Calculate the corrected input power using the following equation co...

Страница 40: ...input frequency 400 MHz Vertical range 1 Vpk pk Sample rate 1 GS s Minimum number of points 1 048 576 samples 3 Configure the calibrator and generate a waveform with the following characteristics Wave...

Страница 41: ...calibrator to measure capacitance 4 Enable the calibrator 5 Wait 1 second for settling then record the measured capacitance 6 Compare the measured capacitance to the appropriate Limit from the Input...

Страница 42: ...1 M 100 V 0 050 272E 6 1 Connect the 50 BNC terminator m to channel 0 of the PXIe 5164 2 Configure the PXIe 5164 with the following settings Input impedance the Input Impedance value from the RMS Noi...

Страница 43: ...g Timebase procedure on channel 0 This procedure only needs to run on one channel 6 Complete the Adjusting 1 M Compensation Attenuator procedure on channel 1 7 Complete the Adjusting 1 M DC Reference...

Страница 44: ...or Cal Adjust VI with the following settings frequency generated Hz The frequency of the square waveform present on the specified channel of the PXIe 5164 amplitude generated Vpk pk The amplitude of t...

Страница 45: ...to generate and configure the calibrator to output the specified DC voltage into 50 4 Enable the calibrator output if not already enabled 5 Wait 1 second for settling 6 Call the niHsai Group B Externa...

Страница 46: ...Return to the main Adjusting the PXIe 5164 task Adjusting 50 Passband Amplitude Flatness and Bandwidth Follow this procedure to adjust the 50 passband amplitude flatness and bandwidth of the PXIe 5164...

Страница 47: ...f Adapter 3 SMA m to BNC m Adapter 4 Power Splitter 5 PXIe 5164 6 SMA m to SMA m Cable 7 SMA f to N m Adapter 8 Signal Generator 2 Call the niHsai Group B External Cal API v1 Host lvclass Passband Fla...

Страница 48: ...the current Test Point Frequency 10 Call the niHsai Group B External Cal API v1 Host lvclass Passband Flatness Cal Adjust VI with the following settings actual frequency generated Hz The frequency of...

Страница 49: ...8 PXIe 5164 400 MHz Oscilloscope ACCESS ACTIVE CH 0 CAT II TO 50 5 V Peak MAX 1 M 250 V Peak MAX CH 1 250 V RMS MAX INPUT 5 V MAX CLK IN PFI 0 AUX 0 2 1 3 4 5 3 1 Power Sensor 2 BNC f to N f Adapter...

Страница 50: ...quency using the power sensor frequency correction function 7 Wait 0 1 second for settling 8 Use the power sensor to measure the power in dBm Record the result as measured input power 9 Calculate the...

Страница 51: ...phone assistance from NI Application Engineers Visit ni com services for NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your NI product...

Страница 52: ...nts EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy an...

Отзывы: