Table 3. DC Accuracy Verification (Continued)
Config
Input
Impedance
(
Ω)
Vertical
Range
(V
pk-pk
)
Vertical
Offset
(V)
Test
Points
(V)
As-
Found
Limits
(mV)
As-Left
Limits
(mV)
Measurement
Uncertainty
(mV)
1
43
1 M
100 V
0
45.00
±493
±392
±34.0
44
1 M
100 V
0
-45.00
±493
±392
±34.0
45
1 M
100 V
50
95.00
±693
±502
±66.0
46
1 M
100 V
-50
-95.00
±693
±502
±66.0
47
1 M
100 V
-155.0
-200.0
±1113
±802
±87.0
48
1 M
50 V
-177.5
-200.0
±956
±658
±72.0
49
1 M
25 V
-188.8
-200.0
±878
±597
±54.0
50
1 M
10 V
-195.5
-200.0
±831
±552
±55.0
51
1 M
5 V
-197.8
-200.0
±816
±540
±48.0
52
1 M
2.5 V
-198.9
-200.0
±808
±534
±48.0
53
1 M
100 V
155 .0
200.0
±1113
±802
±87.0
54
1 M
50 V
177.5
200.0
±956
±658
±72.0
55
1 M
25 V
188.8
200.0
±878
±597
±54.0
56
1 M
10 V
195.5
200.0
±831
±552
±55.0
57
1 M
5 V
197.8
200.0
±816
±540
±48.0
58
1 M
2.5 V
198.9
200.0
±808
±534
±48.0
1.
Connect the calibrator test head to channel 0 of the PXIe-5164.
2.
Configure the PXIe-5164 with the following settings:
•
Input impedance: the Input Impedance value from the
table
•
Maximum input frequency: (1 MΩ) 300 MHz, (50 Ω) 400 MHz
•
Vertical offset: the Vertical Offset value from the
table
•
Vertical range: the Vertical Range value from the
1
Measurement uncertainty based on Fluke 9500B with Fluke 9530 test head specifications that
apply at T
cal
± 5 °C, where Factory T
cal
= 23 °C. Uncertainty of the 9500B includes long-term
stability of 1 year, temperature coefficient, linearity, load, and line regulation and traceability of
factory and National Calibration Standard.
18
|
ni.com
|
PXIe-5164 Calibration Procedure