Chapter 3
Configuring and Testing
©
National Instruments Corporation
3-5
NI-DAQmx
In NI-DAQmx, you can configure software settings such as bridge
configuration, voltage excitation level, filter bandwidth, gain/input range,
and calibration settings in the following two ways:
•
NI-DAQmx Task
or
Global Channel
in MAX
•
Functions in your application
Note
Some software configurable settings can only be set through your application. This
section only discusses settings available in MAX. Refer to Chapter 5,
, for information about using functions in your application.
The following sections describe settings you can change in MAX and
where they are located. Strain and custom voltage with excitation are the
most commonly used
NI-DAQmx Task
or
NI-DAQmx Global Channel
types with the NI PXI-4220. Use the
Custom Voltage with Excitation
NI-DAQmx Task
or
Global Channel
when measuring load, force, torque,
pressure or other bridge-based sensors. You can configure the following
settings using MAX or your application.
•
Bridge Configuration—set through the settings tab of your
NI-DAQmx Task
or
NI-DAQmx Global Channel
and functions
in your application. Virtual channel properties override module
properties. The default bridge configuration for NI-DAQmx is full
bridge.
•
Voltage Excitation— set either through
NI-DAQmx Task
or
NI-DAQmx Global Channel
. You also can set the voltage excitation
level through your application. In NI-DAQmx you can choose from
contiguous voltages between 0 and 10 V. The default voltage
excitation in NI-DAQmx is 0 V.
•
Lowpass Filter Cutoff Frequency—set using the
Device
tab through
either the
NI-DAQmx Task
or
NI-DAQmx Global Channel
.
You also can set it through your application. The default filter cutoff
frequency in NI-DAQmx is 10 kHz.
•
Input Range—set the input range through
NI-DAQmx Task
or
NI-DAQmx Global Channel
. When you set the minimum and
maximum range of the
NI-DAQmx Task
or
NI-DAQmx Global
Channel
, the driver selects the best gain for the measurement. You also
can set it through your application.
•
Calibration Settings—set offset null compensation potentiometer
settings and control shunt calibration switches only through
Strain
NI-DAQmx Task
or
Strain NI-DAQmx Global Channel
or through
your application. The
Custom Voltage with Excitation
NI-DAQmx