Common Questions
This appendix lists common questions related to the use of the
NI PXI-4220.
Which version of NI-DAQ works with the NI PXI-4220 and how do I
get the most current version of NI-DAQ?
You must have NI-DAQ 7.0 or later and use NI-DAQmx. Visit
ni.com
,
select
Download Software»Drivers and Updates»Search Drivers and
Updates
, and enter the keyword
NI-DAQ
to find the latest version of
NI-DAQ for your operating system.
When no signal is connected to the NI PXI-4220, what sort of behavior
should I expect?
While the NI PXI-4220 may react differently due to system and condition
variables, in most cases, when no signal is connected to the NI PXI-4220
the instrumentation amplifiers float, and the signal will likely drift to the
positive or negative voltage rail.
How do I control the gain, excitation voltage, filter setting, bridge
configuration, potentiometer settings, and shunt calibration switches
from LabVIEW?
The gain of each NI PXI-4220 channel is automatically set based on the
channel limits used in setting up the acquisition. Although excitation
voltage, filter setting, and bridge configuration are usually set using MAX,
you also can control or change these settings programmatically using a
DAQmx Channel Property Node
in LabVIEW.
Are there any special considerations for an NI PXI-4220 channel
not
connected to a Wheatstone bridge?
If you are attempting to make a voltage with no bridge signal conditioning,
you should set the excitation voltage to zero and the bridge configuration
to full bridge. This is because the NI PXI-4220 input offset correction
constants are obtained with the excitation set to 0 V. With the excitation
voltage other than zero, error voltages from the auto-nulling circuits can