Figure
Callout
Connection
1
The first mmRH-5582 module, RF0, connects to the PXIe-5831 (not pictured).
2
The second mmRH-5582 module, RF1, connects to the PXIe-5831 (not
pictured).
3
DFF
A 3 dB attenuator connects to the first mmRH-5582 TRX Port 0,
which then connects to the chamber top bulkhead 1A port using an 8 in.
1.85 mm-to-1.85 mm cable.
CATR
A 3 dB attenuator connects to the first mmRH-5582 TRX Port 0,
which then connects to the chamber rear bulkhead 1B port using a 2.92 mm
(m)-to-2.4 mm (f) adapter and 23.6 in. 2.4 mm-to-2.4 mm cable.
4
DFF
A 3 dB attenuator connects to the first mmRH-5582 module TRX Port
1, which then connects to the chamber top bulkhead 2A port using an 8 in.
1.85 mm-to-1.85 mm cable.
CATR
A 3 dB attenuator connects to the first mmRH-5582 module TRX Port
1, which then connects to the chamber rear bulkhead 2B port using a 2.92 mm
(m)-to-2.4 mm (f) adapter and 23.6 in. 2.4 mm-to-2.4 mm cable.
5
A 3 dB attenuator connects to the second mmRH-5582 module TRX Port 0,
which then connects to the chamber rear bulkhead 1A port using a 48 in.
2.4 mm-to-2.4 mm cable.
6
A 3 dB attenuator connects to the second mmRH-5582 module TRX Port 1,
which then connects to the chamber rear bulkhead 2A port using a 48 in.
2.4 mm-to-2.4 mm cable.
7
The PXIe-6361 Connector 0 connects to the chamber control panel VHDCI
connection using the SHC68-C68-RDIO2 digital cable.
8
The PXI chassis Ethernet connects to the chamber control panel Ethernet
connection.
Related Information
on page 5
mmWave OTA Validation Test Reference Architecture
|