
Connecting the Chamber to a Client PC
Connect the mmWave OTA Validation Anechoic Chamber to a client PC using a direct
connection.
Tip
For best results, dedicate one Ethernet port for the communication and connect
the Ethernet cable from that port directly to the client PC to avoid slow hubs or
switches.
Note
The Ethernet port on the client PC must use the same subnet as the host
chamber. For example, if you use the default fixed IP address for the chamber
(
192.168.2.10
), use
192.168.2.1
for the client PC. If you are connecting
multiple clients to the host chamber through a switch, each client must have a unique
IP address on the same subnet as the host.
Verify the Connection
To verify that the chamber is connected to the client PC, open a Command Prompt window
and enter
"ping 192.168.2.10"
.
Configuring Settings in MAX
You must configure and rename your instruments in Measurement & Automation Explorer
(MAX) before you begin taking measurements.
1.
Launch MAX.
2.
In the configuration tree, expand
Devices and Interfaces
to see the list of installed NI
hardware. Installed instruments appear under the name of the associated chassis.
3.
Within the
Settings
section for each instrument, enter the new name in the
Name
textbox
as shown in the following table:
Instrument
Updated Name
PXIe-6361
DAQ
PXIe-3622
mmWave_VST
4.
Click
Save
in the MAX toolbar after you change the name for each instrument.
5.
In the configuration tree, select the PXIe-1095 chassis.
6.
On the
Triggers
tab, route the following PXI trigger lines between the PXIe-6361 and
PXIe-5831:
a)
Set
PXI_Trig0
to be away from the PXIe-6361 bus segment to the PXIe-3622 bus
segment.
b)
Set
PXI_Trig1
to be away from the PXIe-3622 bus segment to the PXIe-6361 bus
segment.
7.
Click
Save
.
18
|
ni.com
|
mmWave OTA Validation Test Reference Architecture