National Instruments Corporation
17
6703/6704 Calibration Procedure
Current Output Calibration
Figure 4.
Current Output Calibration (6704 Only)
AO_VWrite(deviceNumber,16,0.1)
Connect the multimeter to current channel 16.
AO_VWrite(deviceNumber,33,voltage)
Adjust voltage until the DMM reads 0.1 mA.
AO_VWrite(deviceNumber,35,voltage)
Adjust voltage until the DMM reads 20.0 mA.
AO_VWrite(deviceNumber,16,20.0)
Set_DAQ_Device_Info(deviceNumber,ND_CALIBRATION_ENABLE,ND_YES)