National Instruments Corporation
15
6703/6704 Calibration Procedure
Current Output Verification (6704 Only)
Figure 2.
Current Output Verification (6704 Only)
Look at the 6703/6704 specification table to determine the
appropriate ranges for the voltages and currents to be verified.
Get_Cal_Date (deviceNumber)
Connect the multimeter to channel current channel 16.
Verify that the current read by the multimeter is within range.
AO_VWrite(deviceNumber,16,voltage)