6703/6704 Calibration Procedure
16
www.natinst.com
Voltage Output Calibration
Figure 3.
Voltage Output Calibration
Connect the multimeter to channel voltage channel 0.
Set_DAQ_Device_Info(deviceNumber,ND_CALIBRATION_ENABLE,ND_YES)
AO_VWrite(deviceNumber,0,0.0)
AO_VWrite(deviceNumber,0,10.0)
6703:
AO_VWrite(deviceNumber,16,voltage)
Adjust voltage until the DMM reads 10.0 V.
6704:
AO_VWrite(deviceNumber,32,voltage)
Adjust voltage until the DMM reads 10.0 V.
6703:
AO_VWrite(deviceNumber,17,voltage)
Adjust voltage until the DMM reads 10.0 V.
6704:
AO_VWrite(deviceNumber,34,voltage)
Adjust voltage until the DMM reads 10.0 V.