6703/6704 Calibration Procedure
14
www.natinst.com
Flowcharts
These flowcharts show the appropriate NI-DAQ function calls for
verifying and calibrating your 6703/6704 board. Refer to the
section of this document, the
NI-DAQ Function Reference Manual
for PC Compatibles,
and
the NI-DAQ User Manual for PC Compatibles
for
additional information on the software structure.
Voltage Output Verification
Figure 1.
Voltage Output Verification
Look at the 6703/6704 specification table to determine the
appropriate ranges for the voltages and currents to be verified.
Get_Cal_Date (deviceNumber)
Connect the multimeter to channel voltage channel 0.
Verify that the voltage read by the multimeter is within range.
AO_VWrite(deviceNumber,0,voltage)
Get_Cal_Date (deviceNumber)
Verify that the voltage read by the multimeter is within range.
AO_VWrite(deviceNumber,0,voltage)