manualshive.com logo in svg
background image

©

 National Instruments Corporation

9

6703/6704 Calibration Procedure

7.

Again, compare the resulting value shown by the DMM to the upper 
and lower limits listed in 

Specifications

 section table. If the value falls 

between these limits, the board passes current verification.

8.

Disconnect the DMM from the 6704.

You have now verified the current output of your board.

Calibrating Your Board

The 6703/6704 calibration procedure allows you to adjust the output 
calibration constants. At the end of each calibration procedure, these new 
constants will be stored in the user calibration area, ensuring that the board 
stores a permanent record of the calibration constants. Storing the 
calibration constants here also ensures that the constants automatically load 
when the board is powered up.

Calibrating Voltage Output

For the voltage outputs, you need to make offset and gain adjustments. You 
perform offset adjustment first, followed by gain adjustment. You will then 
perform both offset and gain adjustment again to ensure proper accuracy. 
This second set of offset and gain adjustments is necessary because a large 
initial change in one of the calibration channels can affect the accuracy of 
the other calibration channel. The following table lists the offset and gain 
channel numbers.

You need a DMM to measure the voltage outputs. Make sure that you have 
read the 

Equipment and Other Test Requirements

 section listed above, and 

follow these steps:

1.

Run 

Set_DAQ_Device_Info

 with 

infoType

 set to 

ND_CALIBRATION_ENABLE

 and 

infoValue

 set to 

ND_YES

. This step 

configures the board so you can access the calibration channels.

2.

Connect the DMM to the board as shown in the following table.

Calibration 

Channel

6703 Channel 

Numbers

6704 Channel 

Numbers

Voltage Offset

16

32

Voltage Gain

17

34

Signal Name

Description

6703/6704 

Output

DMM Input

VCH0

Voltage Channel 0

34

High

AGND0

Analog Ground 0

68

Low

Содержание 6703

Страница 1: ...it This section also explains internal and external calibration Equipment and Other Test Requirements This section describes what you need to do before you can calibrate your device including an over...

Страница 2: ...of your application indicate how often the 6703 6704 products need to be calibrated to maintain accuracy National Instruments recommends that you perform a complete calibration at least once every ye...

Страница 3: ...the calibration and verification process to approximately 10 minutes However manual calibration and verification can take as long as 1 hour You can automate the adjustment and verification procedure i...

Страница 4: ...e detailed information on using the NI DAQ driver The function reference manual includes information on the functions in the driver The user manual provides instructions on installing and configuring...

Страница 5: ...edure The calibration procedure in the Calibration Process section later in this document provides step by step instructions on calling the appropriate calibration functions These calibration function...

Страница 6: ...as by picking up extra noise that would affect calibration Maintain the temperature between 18 28 C Keep relative humidity below 80 Use shielded copper wire for all cable connections to the device It...

Страница 7: ...his procedure can be easily expanded to verify all the voltage channels by performing the procedure once for each voltage channel To do this perform the procedure once for each value of the channel pa...

Страница 8: ...he procedure once for each value of the channel parameter from 16 to 31 You will also need to connect the appropriate channel to your multimeter each time you repeat the procedure See the PCI PXI 6703...

Страница 9: ...make offset and gain adjustments You perform offset adjustment first followed by gain adjustment You will then perform both offset and gain adjustment again to ensure proper accuracy This second set...

Страница 10: ...the voltage on voltage channel 0 Decreasing the voltage value will increase the magnitude of the voltage on voltage channel 0 You have now calibrated the voltage output Calibrating Current Output 6704...

Страница 11: ...on current channel 16 Decreasing the voltage value will increase the magnitude of the current on current channel 16 You have now calibrated the current output Saving Calibration Values Once you comple...

Страница 12: ...ng tables are the specifications you should use when calibrating and verifying the 6703 6704 The tables show the specifications for 1 year and 24 hour calibration intervals Using the Tables The follow...

Страница 13: ...r Limit V Lower Limit V Upper Limit V 0 0 0 0003125 0 0003125 0 0003125 0 0003125 10 0 9 9992675 10 0007325 9 9990975 10 0009025 Table 2 Current Output Specifications 6704 Only Test Point Value mA 24...

Страница 14: ...r additional information on the software structure Voltage Output Verification Figure 1 Voltage Output Verification Look at the 6703 6704 specification table to determine the appropriate ranges for th...

Страница 15: ...tput Verification 6704 Only Look at the 6703 6704 specification table to determine the appropriate ranges for the voltages and currents to be verified Get_Cal_Date deviceNumber Connect the multimeter...

Страница 16: ...IBRATION_ENABLE ND_YES AO_VWrite deviceNumber 0 0 0 AO_VWrite deviceNumber 0 10 0 6703 AO_VWrite deviceNumber 16 voltage Adjust voltage until the DMM reads 10 0 V 6704 AO_VWrite deviceNumber 32 voltag...

Страница 17: ...6704 Only AO_VWrite deviceNumber 16 0 1 Connect the multimeter to current channel 16 AO_VWrite deviceNumber 33 voltage Adjust voltage until the DMM reads 0 1 mA AO_VWrite deviceNumber 35 voltage Adju...

Страница 18: ...libration Procedure 18 www natinst com Saving Calibration Values Figure 5 Saving Calibration Values Set_DAQ_Device_Info deviceNumber ND_CALIBRATION_ENABLE ND_NO AO_Calibrate deviceNumber 2 1 Copy_Cal...

Отзывы: