manualshive.com logo in svg
background image

6703/6704 Calibration Procedure

2

www.natinst.com

Calibration Overview

This section defines calibration, describes why it is necessary, and explains 
when you should do it. This section also covers internal and external 
calibration.

What Is Calibration?

Calibration is a procedure of reading offset and gain errors from a data 
acquisition (DAQ) board and updating special analog calibration circuitry 
that corrects these errors. National Instruments calibrates every 6703/6704 
product at the factory. During the factory-calibration procedure, the 
calibration constants—values used to update the analog calibration 
circuitry—are stored in nonvolatile memory on the board. From memory, 
these values are loaded and used as needed.

Why Calibrate?

Offset and gain errors may drift with time and temperature. As a result, the 
factory-set calibration constants may become invalid, requiring calibration 
to achieve the specified accuracy of the board.

How Often Should You Calibrate?

The measurement requirements of your application indicate how often the 
6703/6704 products need to be calibrated to maintain accuracy. National 
Instruments recommends that you perform a complete calibration at least 
once every year. This interval may be shortened to 90 days or six months 
depending on the demands of your application. 

Calibration Options: External Versus Internal

6703/6704 products have two different calibration options, an internal or 
self-calibration, and an external calibration. Both of these calibration 
modes remove error from measurements. 

Internal Calibration

All 6703/6704 devices perform continuous internal calibration using 
special calibration channels. These devices each have 16 voltage outputs 
and two voltage calibration channels. The 6704 also has 16 current outputs 
and two current calibration channels. All the output and calibration 
channels get scanned and refreshed together. The outputs of the calibration 
channels for the 6703 and 6704 are continuously compared to the onboard 
reference, and the 16-bit DAC’s offset and gain are adjusted to minimize 
the errors of the calibration channels. Since the 6703 and 6704 have 
excellent channel-to-channel matching of offset and gain errors, 

Содержание 6703

Страница 1: ...it This section also explains internal and external calibration Equipment and Other Test Requirements This section describes what you need to do before you can calibrate your device including an over...

Страница 2: ...of your application indicate how often the 6703 6704 products need to be calibrated to maintain accuracy National Instruments recommends that you perform a complete calibration at least once every ye...

Страница 3: ...the calibration and verification process to approximately 10 minutes However manual calibration and verification can take as long as 1 hour You can automate the adjustment and verification procedure i...

Страница 4: ...e detailed information on using the NI DAQ driver The function reference manual includes information on the functions in the driver The user manual provides instructions on installing and configuring...

Страница 5: ...edure The calibration procedure in the Calibration Process section later in this document provides step by step instructions on calling the appropriate calibration functions These calibration function...

Страница 6: ...as by picking up extra noise that would affect calibration Maintain the temperature between 18 28 C Keep relative humidity below 80 Use shielded copper wire for all cable connections to the device It...

Страница 7: ...his procedure can be easily expanded to verify all the voltage channels by performing the procedure once for each voltage channel To do this perform the procedure once for each value of the channel pa...

Страница 8: ...he procedure once for each value of the channel parameter from 16 to 31 You will also need to connect the appropriate channel to your multimeter each time you repeat the procedure See the PCI PXI 6703...

Страница 9: ...make offset and gain adjustments You perform offset adjustment first followed by gain adjustment You will then perform both offset and gain adjustment again to ensure proper accuracy This second set...

Страница 10: ...the voltage on voltage channel 0 Decreasing the voltage value will increase the magnitude of the voltage on voltage channel 0 You have now calibrated the voltage output Calibrating Current Output 6704...

Страница 11: ...on current channel 16 Decreasing the voltage value will increase the magnitude of the current on current channel 16 You have now calibrated the current output Saving Calibration Values Once you comple...

Страница 12: ...ng tables are the specifications you should use when calibrating and verifying the 6703 6704 The tables show the specifications for 1 year and 24 hour calibration intervals Using the Tables The follow...

Страница 13: ...r Limit V Lower Limit V Upper Limit V 0 0 0 0003125 0 0003125 0 0003125 0 0003125 10 0 9 9992675 10 0007325 9 9990975 10 0009025 Table 2 Current Output Specifications 6704 Only Test Point Value mA 24...

Страница 14: ...r additional information on the software structure Voltage Output Verification Figure 1 Voltage Output Verification Look at the 6703 6704 specification table to determine the appropriate ranges for th...

Страница 15: ...tput Verification 6704 Only Look at the 6703 6704 specification table to determine the appropriate ranges for the voltages and currents to be verified Get_Cal_Date deviceNumber Connect the multimeter...

Страница 16: ...IBRATION_ENABLE ND_YES AO_VWrite deviceNumber 0 0 0 AO_VWrite deviceNumber 0 10 0 6703 AO_VWrite deviceNumber 16 voltage Adjust voltage until the DMM reads 10 0 V 6704 AO_VWrite deviceNumber 32 voltag...

Страница 17: ...6704 Only AO_VWrite deviceNumber 16 0 1 Connect the multimeter to current channel 16 AO_VWrite deviceNumber 33 voltage Adjust voltage until the DMM reads 0 1 mA AO_VWrite deviceNumber 35 voltage Adju...

Страница 18: ...libration Procedure 18 www natinst com Saving Calibration Values Figure 5 Saving Calibration Values Set_DAQ_Device_Info deviceNumber ND_CALIBRATION_ENABLE ND_NO AO_Calibrate deviceNumber 2 1 Copy_Cal...

Отзывы: