Sweep Mode Operating Instructions
4-7
4.5 Marker
Selecting this function a
Marker
is immediately enabled, and it appears on
the screen as a small black pointing down arrow corresponding to the
highest reading; simultaneously a small window shows up in the bottom left
corner of the screen, indicating the actual frequency and level read by the
marker.
The Marker function is not a simple search for the highest readings on the
screen – it would be a useless exercise that would display a number of
points all grouped together – but it’s a true peak search that measures the
degree of variation of the signal: a peak is so classified only if it “pops-up”
from the adjacent signals with at least 5 dB difference in amplitude.
Following this criterion, if the measured signal is a flat line no peaks will be
found.
Selecting
Highest X
– where X represents the maximum number of markers
found by the receiver during its sweep – few more markers are added, each
one “diamond” shaped, defined as per the previous explanation.
The maximum number of markers available is 10.
By using arrow keys it is possible to quickly move on markers from higher
amplitude (left arrow) to lower amplitude (right arrow), i.e. if the highest is
selected, pressing the right arrow key the 2
nd
highest is selected, then
pressing again the same key the 3
rd
highest and so on and so forth.
It is also possible to use the rotary knob to move from a marker to the next
one.
The marker under analysis (main marker) is a black arrow pointing down,
the others are diamond shaped.
Pressing the
Peak
button the marker returns to the highest reading.
In order to make easier and faster the evaluation of the more noisy signals,
the User can now select either the
Analyzer
or the
Tune
function, entering
respectively into a spectrum analysis of main marker position or in the
manual mode evaluation of it. To operate this two conditions please refer to
the appropriate section of this Operating Manual.
In manual mode, i.e. having selected
Tune
on the main marker, it is possible
to navigate from one peak to the other just pressing the arrow keys, without
to come back to the Sweep display and remaining in manual mode.
This is another unique feature of PMM 9010 to improve the productivity of
the test lab and make easier the work of the test engineer.
The
Marker Off
exits from the marker function.
Содержание PMM 9010 Series
Страница 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Страница 42: ...2 10 Installation This page has been left blank intentionally ...
Страница 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Страница 72: ...7 6 Applications This page has been left blank intentionally ...
Страница 92: ...9 14 General Information This page has been left blank intentionally ...
Страница 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Страница 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Страница 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Страница 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Страница 230: ...14 36 Remote control This page has been intentionally left blank ...
Страница 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...