Remote control
14-19
SSFD
FreqStart;
FreqStop;
FreqStep;
Detector;
HoldTime;
Rbw;
MinAtt;
Preamp;
Preselector;
ScanHoldT
Note: Please look at §14.7 for further details related to FFT based units (i.e.
9010F)
Arguments are as follows:
•
FreqStart
= Sweep Start Frequency expressed in Hz
•
FreqStop
= Sweep Stop Frequency expressed in Hz
•
FreqStep
= Sweep Step Frequency expressed in Hz
o
If
FreqStep > 0
then a standard sweep, tuned step by step, is
made
o
If
FreqStep < 0
(typical = -1) then the step will be set to AUTO
o
If
FreqStep = 0
then the sweep will be made by tuning all the
frequencies which are listed in Frequency-Scan-Tab (which it
must have been written previously by command
SSFW
.
•
Detector
. In
non –smart mode
, any combination of the following
detectors can be used and combined. All the detectors set will be
measured and sent out. However, in
Smart Mode
only one detector at
time, in addition to PEAK which is always present, can be set.
o
Non Smart Mode
(All detectors can be indicated at once). Order is
not important.
P
Peak
Q Peak
N C-Rms
A
Avg
R
Rms
C
C-Avg
o
Smart Mode
(No more than 2 detectors can be indicated at once)
S
Smart Mode
(this must be the first in the string)
Q Peak
N C-Rms
A
Avg
R
Rms
C
C-Avg
•
HoldTime
= Hold Time for each step and it is expressed in ms. Note that
in case of Smart Mode HoldTime is used only when the disturbance is
measured again by the alternative detector as minimum time, depending
on the Rbw set, is used for the fast scan.
•
Rbw
. This is the index of Rbw to be used for the scan. Neither automatic
selection, nor automatic switch is performed. The user must issue a
suitable Rbw index. Index are as follows:
o
1
300kHz
o
2
100kHz
o
3
30kHz
o
4
10kHz
o
5
3kHz
o
Only for 9010
6
9kHz (Cispr 16-1)
7
200Hz (Cispr 16-1)
11
10kHz (MIL-STD-461) (only when available)
22
1kHz (MIL-STD-461) (only when available)
23
100Hz (MIL-STD-461) (only when available)
24
10Hz (MIL-STD-461) (only when available)
o
Only for Downconverter
9
1MHz (Cispr 16-1)
10
120kHz (Cispr 16-1)
12
9kHz (Cispr 16-1); No QPeak available
Содержание PMM 9010 Series
Страница 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Страница 42: ...2 10 Installation This page has been left blank intentionally ...
Страница 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Страница 72: ...7 6 Applications This page has been left blank intentionally ...
Страница 92: ...9 14 General Information This page has been left blank intentionally ...
Страница 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Страница 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Страница 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Страница 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Страница 230: ...14 36 Remote control This page has been intentionally left blank ...
Страница 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...