PMM 9030/9060/9180 Frequency Extension
11-51
11.17.3.1 Input:
Attenuator and
preamplifier
Being entered in the Input submenu, to increase or decrease the
attenuation at the input, press
Att +
or
Att -
, and for each touch the
attenuation is increased or decreased of 5 dB (preset value) up to the max
attenuation available of 55 dB.
Depressing either one of these keys force the receiver in manual
attenuation.
The
Min Att
button acts as a toggle switch: it selects or deselects the
minimum attenuation of 10 dB. When the minimum attenuation is selected,
the attenuator – doesn’t matter if in automatic or in manual condition -
cannot be lowered under 10 dB.
Unless specifically required by the test conditions, do not remove the
minimum attenuation of 10 dB.
With the
Preamp
key it is possible to insert or exclude the built-in low noise
preamplification of 10 dB.
The internal 10 dB preamplifier can be used when very weak signals have
to be investigated. As already mentioned, with the preamplifier ON the
receiver takes automatically care of the 10 dB gain when measuring the
signals.
The Preamplifier must be mandatory switched off (default) while carrying
on CISPR measurements.
The
Att Auto
button is used to switch from the two conditions of manual or
automatic attenuation setting.
Please note that the switching of the attenuators is relatively noisy and you
can perceive it distinctly with a “click” for each switch operation.
Using 0 dB attenuation PMM 9030/9060/9180 has no input protection.
This is a potentially dangerous condition for the input stage of the
receiver.
Use 0 dB attenuation only if you are very sure that your input signal is
less than 1 V (or 120 dB
μ
V).
Before to apply an unknown signal to PMM 9030/9060/9180 receiver,
use an oscilloscope or a wide band RF voltmeter to measure it. In any
case set Min. ATT at 10 dB and select the maximum available
attenuation with preamplifier OFF.
If needed, add an external coaxial attenuator on the input signal line.
In case an OVERLOAD indication occurs it is necessary to add an
external attenuator to carry on the measurement without overpassing
the declared limit for continuous power or spectral density.
Содержание PMM 9010 Series
Страница 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Страница 42: ...2 10 Installation This page has been left blank intentionally ...
Страница 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Страница 72: ...7 6 Applications This page has been left blank intentionally ...
Страница 92: ...9 14 General Information This page has been left blank intentionally ...
Страница 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Страница 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Страница 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Страница 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Страница 230: ...14 36 Remote control This page has been intentionally left blank ...
Страница 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...