Remote control
14-21
(Continued)
SSFD
FreqStart;
FreqStop;
FreqStep;
Detector;
HoldTime;
Rbw;
MinAtt;
Preamp;
Preselector;
ScanHoldT
•
SFD=ERR 4
<RC><LF> An Error was found in
HoldTime
value.
HoldTime
is either too big (>10s) or negative.
•
SFD=ERR 5
<RC><LF> An Error was found in
RBW
. Errors can be one
of the following:
o
Unsuitable RBW for the active unit (for instance 200HZ Rbw
while Downconverter is active)
o
Unsuitable RBW for the detector previously selected (for
instance 100kHZ Rbw and QuasiPeak selected).
•
SFD=ERR 6
<RC><LF> An Error was found in
MinAtt
value. Errors can
be one of the following:
o
MinAtt
value is negative
o
MinAtt
value is higher than maximum available attenuation
o
MinAtt
value is not in step of 5
•
SFD=ERR 7
<RC><LF> An Error was found in
Preamp
string.
Preamp
string can be either
ON
or
OFF
only. String is not case sensitive.
•
SFD=ERR 8
<RC><LF> An Error was found in
Preselector
string.
Preselector
string can be either
ON
or
OFF
only. String is not case
sensitive.
Once the sweep is started, the 9010 sends the value of each detector chosen in
the command. Each detector is made of a Little Endian 16 bit integer which
represents the value of that particularly tuned frequency. The value is expressed
in
hundredth of dBm
. Therefore, for each step, there will be a packet of
n
detectors (
n
*2 bytes).
The order of detectors is fixed and is not related to the order issued in the
command. For instance, PAQ is equivalent to QAP as well as RA is same as AR.
The order of detector is as follows (from the first to the last):
1. Peak (which is always present)
2. QPeak
(QuasiPeak)
3. RMS
4. AVG
(Average)
5. C-RMS
6. C-AVG
Special meaning has the value
NOLEVEL
,
-32700
(Hex 8044), which informs
that the detector has not been measured. This happens, for example, in Smart
mode where the alternative detector is taken only for those frequencies the Peak
was above the limit.
Содержание PMM 9010 Series
Страница 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Страница 42: ...2 10 Installation This page has been left blank intentionally ...
Страница 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Страница 72: ...7 6 Applications This page has been left blank intentionally ...
Страница 92: ...9 14 General Information This page has been left blank intentionally ...
Страница 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Страница 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Страница 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Страница 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Страница 230: ...14 36 Remote control This page has been intentionally left blank ...
Страница 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...