14-26
Remote control
14.6.3 Analyzer
Reply
The PMM 9010 replies to the command “
SAGO
” sending back an array of bytes
which contains all the information needed to draw a sweep.
Typically, the user should first send the PMM 9010 all the setting commands to
insure the receiver is correctly set on the wanted parameters and then read the
reply.
Just as an example, the user could send the following commands:
14.6.3.1 Reply
example
#SAAT -1*
Set Attenuator Automatic mode
#SAPA Off*
Switch Preamplifier OFF
#SPLM a*
Switch Pulse Limiter OFF
#SAFF
150e3,30e6*
Sets the Start frequency to 150kHz
and Stop frequency to 30MHz
#SAHT 2*
Sets the hold time to 2 ms
#SADT 2*
Sets the detector to AVG
............. More
Command
............. More
Command
#SAGO *
Start Sweep in Analyzer Mode
The PMM9010 replies by sending a 48 bytes header followed by a variable size
array of integer (the size depends on frequency settings) as follows:
HEADER
ID Byte Value
Meaning
1
A
The first 8 Bytes are just a fixed header and acknowledge the command
2 G
3 O
4 =
5 O
6 K
7 0x13
(CR)
8 0x10
(LF)
9 Start
Frequency
Little Endian 32 bit Float figure representing the start frequency of sweep
10
11
12
13 Stop
Frequency
Little Endian 32 bit Float figure representing the Stop frequency of sweep
14
15
16
17 Sep
Frequency
Little Endian 32 bit Float figure representing the step frequency of sweep
18
19
20
21
…
26
Reserved
…
…
27
Attenuator
Little Endian 16 bit integer representing the attenuator used
28
29
…
48
Reserved
….
…
Содержание PMM 9010 Series
Страница 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Страница 42: ...2 10 Installation This page has been left blank intentionally ...
Страница 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Страница 72: ...7 6 Applications This page has been left blank intentionally ...
Страница 92: ...9 14 General Information This page has been left blank intentionally ...
Страница 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Страница 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Страница 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Страница 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Страница 230: ...14 36 Remote control This page has been intentionally left blank ...
Страница 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...