Remote control
14-25
STGL l
This setting command sets the output level of 9010 Built-In Generator. The string (l)
should be expressed in dBµV.
Level should range between 60 and 90 inclusive
The reply is
TGL=OK
which acknowledges the command has been granted or
TGL
=SERR
if the command has been ignored.
Example:
# STGL 85.5*
Sets the output level of 9010 Built-In Generator to
85.5dBµV.
STGS a
This setting command switches the 9010 Built-In Generator; according the
parameter (a) which can be either
On
or
OFF
.
The reply is
TGS
=OK
which acknowledges the command has been granted or
TGS
=SERR
if the command has been ignored.
Example:
# TGS
ON*
Example:
# TGS
OFF*
STGT a
This setting command changes 9010 Built-In Generator tracking mode; according
the parameter (a) which can be either
On
or
OFF
.
When tracking mode is ON the 9010 Built-In Generator is tuned on the same
receiver frequency. When tracking mode is OFF the 9010 Built-In Generator is
tuned frequency previously set by command STGF
The reply is
TGT =OK
which acknowledges the command has been granted or
TGT
=SERR
if the command has been ignored.
Example:
# TGT ON*
Example:
# TGT OFF*
SUPP n
This setting command
#SUPP n*
outputs on the User Port bit by bit the argument n .
Each Pin has its own contribution according to its index as follows:
OUT0= 2^0
OUT1= 2^1
OUT2= 2^2
OUT3= 2^3
SupplyOFF=2^4
Thus, the argument sent is the sum of each weighted contribution
The reply is
UPP=OK
which acknowledges the command has been granted or
UPP=SERR
if the command has been ignored (IE argument >31).
Example
: SUPP 5
will set:
•
user port pins OUT0 and OUT2 to high level
•
user port pins OUT1 and OUT3 are low level
•
User Port Power Supply (+12V) ON
Содержание PMM 9010 Series
Страница 24: ...XXIV Safety considerations This page has been left blank intentionally ...
Страница 42: ...2 10 Installation This page has been left blank intentionally ...
Страница 66: ...6 6 Manual Mode Operating Instructions This page has been left blank intentionally ...
Страница 72: ...7 6 Applications This page has been left blank intentionally ...
Страница 92: ...9 14 General Information This page has been left blank intentionally ...
Страница 112: ...10 20 Click Mode Operating Instructions This page has been left blank intentionally ...
Страница 145: ...PMM 9030 9060 9180 Frequency Extension 11 33 Screwing the N N Adapter to BL 01 connector ...
Страница 170: ...11 58 PMM 9030 9060 9180 Frequency Extension This page has been left blank intentionally ...
Страница 178: ...12 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally ...
Страница 194: ...13 16 APD Mode Operating Instructions This page has been intentionally left blank ...
Страница 230: ...14 36 Remote control This page has been intentionally left blank ...
Страница 284: ...Suggerimenti Commenti Note Suggestions Comments Note ...