1-32
General information
ELECTRIC FIELD PROBE EP-105
Frequency range
100 kHz - 1000 MHz
Level range
0.05 - 50 V/m
Overload
> 100 V/m
Dynamic range
> 60 dB
Resolution
0.01 V/m
Sensitivity
0.05 V/m
Absolute error @ 50 MHz 6 V/m
±
0.8 dB
Flatness (10 - 300 MHz)
±
0.5 dB
Flatness (300 kHz - 1 GHz)
±
1 dB
Isotropicity
±
0.8 dB (typical
±
0.5 dB @ 930 MHz)
H-field rejection
> 20 dB
Temperature error
0.05 dB/°C
Calibration
Internal into E
2
PROM
Size
350 mm length, 133 mm diameter
Weight
290 g
Fig. 1-25
EP-105 probe
Typical frequency response for EP-105 probe
-8,0
-7,0
-6,0
-5,0
-4,0
-3,0
-2,0
-1,0
0,0
1,0
0,1
1
10
100
1000
10000
[MHz]
[dB]
Содержание LR-01
Страница 72: ...2 12 Installation and use This page has been left blank intentionally...
Страница 76: ...3 4 Probes Manager software installation Click Next to proceed installing...
Страница 116: ...5 4 Firmware Update This page has been intentionally left blank...
Страница 168: ...9 40 Command protocol This page has been left blank intentionally...
Страница 192: ...11 4 Service This page has been left blank intentionally...
Страница 194: ...Suggerimenti Commenti Note Suggestions Comments Note...