9-2
Command protocol
At power ON, LR-01 is automatically set to listen to both the USB and the
optic port for incoming commands.
The optic port will be always available for communicating with the unit.
User’s own software can ask for measurement data when desired or can
switch the LR-01 to master mode where measurement data are
continuously sent over the communication port, without the need of asking
for each.
At any time the operation mode can be switched back to slave mode.
While operating in slave mode, LR-01 sends answers to the received
commands, according to the communication protocol described below.
The LR-01 automatically turns off 30 minutes after receiving the last
command, to save battery.
Please, do not try any commands not covered in this manual, as
random poking can cause the system to crash or lose data and
calibrations.