General information
1-19
DUAL-BAND ELECTRIC AND MAGNETIC SHAPED FIELD PROBE EHP-2B-01
For ICNIRP 1998 and SC6 2015 *
Electric Field
Magnetic Field
Frequency range
ICNIRP 1998
Occupational
0.5 – 9250 MHz
20 – 1000 MHz
General Public
3 – 9250 MHz
SC6 2015
Controlled
Uncontrolled
Level range
(1)
Occupational / Controlled
0.1 – 1000 %
0.3 – 1000 %
General Public / Uncontrolled
0.5 – 1000 %
1.5 – 1000 %
Overload
2000 %
Linearity
(2)
+/- 0.5 dB
Power (Amplitude)
dynamic range
Occupational / Controlled
40 (80) dB
35 (70) dB
General Public / Uncontrolled
33 (66) dB
28 (56) dB
Resolution
0.01 %
Sensitivity
Occupational / Controlled
0.1 %
0.3 %
General Public / Uncontrolled
0.5 %
1.5 %
Frequency flatness
(3)
(typ)
ICNIRP 1998
Occupational
0.5 – 3 MHz +4/-2 dB
3 – 9250 MHz +/-3 dB
20 – 1000 MHz +/-3 dB
General Public
3 – 10 MHz +2/-3 dB
10 – 9250 MHz +/-3 dB
SC6 2015
Controlled
3 – 9250 MHz +/-3.5 dB
Uncontrolled
Anisotropy
(4)
+/- 0.5 dB
Temperature error
(4)
0.03 dB/°C
0.01 dB/°C
Temperature sensor
On board
Field sensor
Triaxial orthogonal dipoles Triaxial orthogonal loops
A/D convertion
On board
Calibration
(5)
internal E
2
PROM
Operating temperature
-20 to +55 °C
Operating relative humidity
(6)
5 to 95 %
Storage temperature
-30 to +75°C
Dimensions
450 mm length, 55mm diameter
Weight
200 g
Specification are subject to change without notice
When not differently specified the following specifications are referred to operating ambient temperature 23°C and relative humidity 50%.
Note (1): Power density referred.
Note (2): At 50 MHz on related level range 6dB above the noise floor
Note (3): Relative to 10% of the standard limit
Note (4): At 50 MHz / 10% of the standard limit
Note (5): Recommended re-calibration interval 24 month
Note (6): Without condensation
(*) All probes include on board A/D conversion, calibration factors on E
2
PROM, and temperature sensor
Fig. 1-14
EHP-2B-01 probe
Содержание LR-01
Страница 72: ...2 12 Installation and use This page has been left blank intentionally...
Страница 76: ...3 4 Probes Manager software installation Click Next to proceed installing...
Страница 116: ...5 4 Firmware Update This page has been intentionally left blank...
Страница 168: ...9 40 Command protocol This page has been left blank intentionally...
Страница 192: ...11 4 Service This page has been left blank intentionally...
Страница 194: ...Suggerimenti Commenti Note Suggestions Comments Note...